MOL2150 Simplexa HSV 1&2 Direct recallZ-2547-2020 · DiaSorin Molecular LLC · initiated May 12, 2020
Class II Terminated All recalls posted in July 2020
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from DiaSorin Molecular LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Due to a software malfunction in the Multi Assay Suite, there is the probability of increased occurrence of false negative results.
FDA root cause category: Under Investigation by firm
What the company is doing
On 05/12/2020, the firm sent an "URGENT FIELD SAFETY NOTIFICATION LETTER" with an acknowledgement form to customers and distributors via email to inform them of an internal observed issue associated with the use of their Multi-Assay Suite (MAS) featured in the software version 2.0 where it may incorrectly apply the thresholds of the first loaded assay to the analytes coupled with the same dye in the second assay resulting in false negative or false positive results. The firm is instructing customers to: -Immediately discontinue the use of the MAS feature and that the issue will be correct in the new Software Version 2.1 which is expected to be release in June 2020. -Fill out the Response Form and return by either: -fax (562) 240-6526 Attention: Technical Services -email to: technical [email protected] -mail to: ATTN: Technical Services, 11331 Valley View Street, Cypress, CA 90630 -Forward this communication to all required individuals with their organization. For further assistance with evaluating assay runs or for any questions, to contact Technical Services at +1-5672-240-6500, option 3 or via email [email protected]
Recall details
- Recall number
- Z-2547-2020
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- May 12, 2020
- Posted by the FDA
- Jul 10, 2020
- Recalling firm
- DiaSorin Molecular LLC
- Quantity
- 13943 plastic vials (24 vials per box)
- Product code
- OQO
- FDA submissions
- K141458, K150962
- Distribution
- Worldwide distribution - U.S. Nationwide distribution including in the states of CO, FL, KS, MI, NC, PA, TX, and WA The countries of Austria, Czech Republic, France and Switerland.
Product description
MOL2150 Simplexa HSV 1&2 Direct -CSF -Genital Swab -Cutaneous / Mucocutaneous Swabs Product Usage: is intended for use on the LIAISON MDX instrument for the qualitative detection and differentiation of HSV-1 and HSV-2 present in cerebrospinal fluid (CSF), or cutaneous and mucocutaneous lesion swabs from patients with signs and symptoms of HSV-1 or HSV-2 infection of the central nervous system (CNS) or cutaneous and mucocutaneous lesions.
Code information (models, lots, serial numbers)
Show the full code information (2k characters)
MOL2150 Simplexa HSV 1&2 Direct: Lot Numbers Release Date Exp Date 3606N 08/16/18 6/30/2020 3607N 08/31/18 6/30/2020 3609N 08/31/18 6/30/2020 3610N 10/08/18 7/31/2020 3615N 01/07/19 6/30/2020 3616N 11/28/18 6/30/2020 3618N 12/26/18 6/30/2020 3619N 12/26/18 6/30/2020 3847N 12/31/18 6/30/2020 3998N 07/20/18 5/31/2020 4000N 07/18/18 5/31/2020 4130N 03/05/19 6/30/2020 4131N 02/01/19 6/30/2020 4132N 02/15/19 6/30/2020 4134N 04/18/19 8/31/2020 4137N 04/04/19 8/30/2020 4138N 04/30/19 8/31/2020 4139N 07/12/19 5/31/2021 5171N 09/17/19 6/30/2021 5172N 09/03/19 6/30/2021 5173N 05/03/19 8/31/2020 5174N 06/04/19 8/31/2020 5175N 06/04/19 8/31/2020 5187N 07/19/19 5/31/2021 5188N 06/05/19 8/31/2020 5189N 07/25/19 5/31/2021 5190N 05/23/19 8/31/2020 5623N 07/25/19 5/31/2021 5738N 08/02/19 5/31/2021 5739N 08/20/19 6/30/2021 5793N 10/07/19 7/31/2021 5795N 03/09/20 12/31/2021 5841N 10/22/19 8/31/2021 5843N 10/31/19 8/31/2021 5845N 10/18/19 8/31/2021 5846N 04/06/20 1/31/2022 5846NA 04/16/20 1/31/2022 5847N 12/17/19 10/31/2021 5878N 03/11/20 1/31/2022 5888N 03/18/20 1/31/2022 R5170N 06/20/19 8/31/2020 R5201N 06/28/19 8/31/2020 R5202N 07/03/19 4/30/2021 R5283N 08/13/19 8/31/2020 R5285N 06/20/19 8/31/2020
Other devices under the same FDA submission
1 device cleared under K141458, K150962 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Simplexa™ HSV 1 & 2 Direct MOL2150Diasorin Molecular LLC · Hsv-1 And Hsv-2 Cns Nucleic-Acid Based Panel1 death1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page