MEVION S250i, MEVION S250 recallZ-2406-2020 · Mevion Medical Systems, Inc. · initiated May 15, 2020

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Mevion Medical Systems, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Couch Correction moves sent after using a 3D CT scan are partially lost if the previous correction sent or recorded had some of the Couch Correction checkboxes turned off

FDA root cause category: Software design

What the company is doing

Mevion issued letter (pCAR0012) on May 15, 2020 via mail or FedEx, with return receipt confirmation. Letter states reason for recall, health risk and action to take: Until a software fix is applied for this problem, the user is cautioned to carefully pay attention to the 3 indicators of Couch corrections. If they disagree with the intended moves the user may simply repeat the alignment step and the problem should clear. Required User Action When using multimodal imaging the user should always check the indicators for Couch Corrections before they are applied. This includes the Verity Couch Correction widget, the Verity dialog box the informs the user of the actions being sent to the Couch, and the In Room Monitor lower right panel that displays the moves before they are made. If these disagree with the intended moves, the user should back up a step and resend the corrections. Users must confirm positioning for every applied Couch correction that all intended moves are indicated and made.

Recall details

Recall number
Z-2406-2020
Status
Terminated
Initiated by the company
May 15, 2020
Recalling firm
Mevion Medical Systems, Inc.
Quantity
8 units
Product code
LHN
FDA submissions
K120676, K172848
Distribution
Distribution to US in MO, NJ, OH, OK, FL, and DC, and the Netherlands

Product description

MEVION S250i, MEVION S250; Proton Radiation Treatment System

Code information (models, lots, serial numbers)

Serial Numbers: MEVION S250-0001 MEVION S250-0002 MEVION S250-0003 MEVION S250-0004 MEVION S250-0005 MEVION S250-0006 MEVION S250-0007 MEVION S250-0008  UDI: (01)00864366000100 (11) 131209 (01)00864366000100 (11)150526 (01)00864366000100 (11)150113 (01)00864366000100 (11)150430 (01)00864366000100 (11)160127 (01)00864366000100 (11)160603 (01)00864366000124 (11) 180220 (01)00864366000124 (11) 181219