Computed Tomography X-ray System, Model: uCT 550, GTIN:… recallZ-2641-2020 · Shanghai United Imaging Healthcare Co., Ltd. · initiated May 23, 2020

Class II Open, Classified All recalls posted in July 2020

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Shanghai United Imaging Healthcare Co., Ltd. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Two issues were identified with the computed tomography x-ray system including a service function which may cause false marking of a bad channel resulting in ring artifacts, and potential intermittent scout scanning interruption due to occasional angle signal drift.. If these problems occur, it may be necessary to rescan the patient resulting in an additional dose of radiation and the possible need for additional contrast medium.

FDA root cause category: Software design

What the company is doing

On 05/23/2020, Customer Notification Letters dated 5/22/2020 were emailed to customers. Customers were asked to do the following: For safety issue 1: If you observe image artifacts, please reboot the system and continue the reconstruction. If the image artifacts persist, please contact the recalling firm's service support team or call the Customer Contact Center at (855) 221-1552 (toll free) or (832) 699-2799 (direct). For safety issue 2: If the scout scanning is interrupted, try repeating the scout scanning. If problem persists, please contact the recalling firm's service support team or call the Customer Contact Center at (855) 221-1552 (toll free) or (832) 699-2799 (direct). Customers were asked to notify all users of the affected products via Customer Notification Letter. On 8/30/2024, a correction notice was sent identifying an additional serial number.

Recall details

Recall number
Z-2641-2020
Recall class
Class II
Status
Open, Classified
Initiated by the company
May 23, 2020
Posted by the FDA
Jul 20, 2020
Recalling firm
Shanghai United Imaging Healthcare Co., Ltd.
Quantity
3
Product code
JAK
FDA submissions
K181414
Distribution
US Distribution to state of: TX

Product description

Computed Tomography X-ray System, Model: uCT 550, GTIN: 06971576831036 Intended use The uCT Computed Tomography X-ray System uCT 530/uCT 550 is a computed tomography X-ray system intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes and indicated for the whole body (including head, neck and vascular).

Code information (models, lots, serial numbers)

Serial Numbers: 353020, 353021, and 353023 with software version R001.11.0.729741-Re-1100