uCT recallZ-2472-2020 · Shanghai United Imaging Healthcare Co., Ltd. · initiated May 8, 2020

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Shanghai United Imaging Healthcare Co., Ltd. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Two issues were identified with the computed tomography x-ray system including a service function which may cause false marking of a bad channel resulting in ring artifacts, and potential intermittent scout scanning interruption due to occasional angle signal drift.. If these problems occur, it may be necessary to rescan the patient resulting in an additional dose of radiation and the possible need for additional contrast medium.

FDA root cause category: Software design

What the company is doing

The firm sent letters to their customers by email on 5-22-2020. The letter informs customers of the two potential safety issues with the device, and notes that these problems may cause image artifacts and scanning workflow interruptions. Risks to patient health are noted to include the possible need to rescan the patient resulting in an additional radiation dose and the possible need for additional contrast medium. Customers were informed that United Imaging has released a field update for the uCT 530 to address the issues. If customers observe image artifacts, they are requested to reboot the system and continue the reconstruction. If the scout scanning is interrupted, customers are instructed to try repeating the scout scanning. If problems persist, customers should contact their United Imaging service support team or call the United Imaging Customer Contact Center at (855) 221-1552 (toll free) or (832) 699-2799 (direct).

Recall details

Recall number
Z-2472-2020
Status
Open, Classified
Initiated by the company
May 8, 2020
Recalling firm
Shanghai United Imaging Healthcare Co., Ltd.
Quantity
4 units
Product code
JAK
FDA submissions
K181414
Distribution
Distributed to Texas.

Product description

uCT 530 Computed Tomography X-Ray System

Code information (models, lots, serial numbers)

Units with software version no. R001.11.0.729741-Re-1100 including the following: Serial number/GTIN: 300162 (GTIN: 6971576831012); 300163 (GTIN: 6971576831012);  300174 (GTIN: 6971576831012); and 300176 (GTIN: 6971576831012).