Skip to content

ADVIA Chemistry Calibrator recallZ-2430-2020 · Siemens Healthcare Diagnostics, Inc. · initiated Jun 5, 2020

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Siemens Healthcare Diagnostics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Calibration errors have been observed with the Total and Direct Bilirubin assays when using Chemistry Calibrator lot 534177 for ADVIA Chemistry and lot 534179 for Atellica CH.

FDA root cause category: Nonconforming Material/Component

What the company is doing

Siemens Healthcare Diagnostics sent a letter notification sent to customers dated June 10, 2020. As one of the short-term alternatives for customers, where regulations permit, Siemens has evaluated the use of the lyophilized RANDOX Calibration Serum Level 3 Lot 1024UE. (CAL 3) for generating DBIL_2/DBil_2 and TBIL_2/TBil_2 results until a new lot of unaffected Siemens Chemistry Calibrator is available.

Recall details

Recall number
Z-2430-2020
Status
Terminated
Initiated by the company
Jun 5, 2020
Recalling firm
Siemens Healthcare Diagnostics, Inc.
Quantity
1872 units
Product code
NDW
FDA submissions
K050374
Distribution
US Nationwide Distribution

Product description

ADVIA Chemistry Calibrator

Code information (models, lots, serial numbers)

Lots: 534177 (including sublots A, B and C) 960742  UDI: (01)00630414223414(10)534177(17)20210930 (01)00630414223414(10)534177A(17)20211031 (01)00630414223414(10)534177B(17)20211231 (01)00630414223414(10)534177C(17)20220131 (01)00630414223414(10)534177D(17)20220331 (01)00630414223414(10)960742(17)20220531

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

2 devices cleared under K050374 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 85844).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page