Skip to content

Computed Tomography X-ray System recallZ-2705-2020 · Shanghai United Imaging Healthcare Co., Ltd. · initiated Jun 20, 2020

Class II Open, Classified All recalls posted in July 2020

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Shanghai United Imaging Healthcare Co., Ltd. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The metal edge overlap of the mylar strip in the gantry was uneven, which may create a sharp edge.

FDA root cause category: Device Design

What the company is doing

A Customer Notification Letter was sent to the affected consignees on 06/20/2020 via email. United Imaging will replace the mylar strip with an improved one on the affected products via on-site update. On 8/30/2024, a correction notice was sent identifying an additional serial number.

Recall details

Recall number
Z-2705-2020
Recall class
Class II
Status
Open, Classified
Initiated by the company
Jun 20, 2020
Posted by the FDA
Jul 27, 2020
Recalling firm
Shanghai United Imaging Healthcare Co., Ltd.
Quantity
3
Product code
JAK
FDA submissions
K181414
Distribution
US Nationwide distribution including in the states of CA, IN and TX. See uploaded distribution list for more details.

Product description

Computed Tomography X-ray System; Model uCT 550; Rx; UDI: (01) GTIN: 06971576831036 - Product Usage: intended to produce cross-sectional images of the body using computer reconstruction of x-ray transmission data taken at different angles and planes and indicated for the whole body (including head, neck, cardiac, and vascular).

Code information (models, lots, serial numbers)

SN 353020, 353021, 353023

Other devices under the same FDA submission

1 device cleared under K181414 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 85868).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page