Verigene Processor SP with assay: Luminex, VERIGENE… recallZ-2775-2020 · Luminex Corporation · initiated Jun 15, 2020
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Luminex Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
An instrument failure mode that may result in a Blood Culture Gram Positive (BC-GP) or Gram Negative (BC-GN) false negative call.
FDA root cause category: Device Design
What the company is doing
On 06/10/2020, the firm sent a letter to customers titled, "MEDICAL DEVICE FIELD ACTION", regarding an instrument failure mode that may result in a Blood Culture Gram Positive (BC-GP) or Gram Negative (BC-GN) false negative call. The firm has developed a software solution that identifies and alerts users to the issue. The firm will contact customers to verify their address to ship the software material and provide installation information. Once the update has been completed, customers need to email a screen shot of the updated software, as well as a completed Notice of Destruction to Global Support Services at [email protected] to complete the upgrade process. The firm is also stating that until the updated software is installed, it is important, and part of good laboratory practice, that the customer follow the specific guidelines indicated in the current package inserts. Specifically, "Not Detected" test results from a positive blood culture should be evaluated further. The sub-culturing of blood cultures is necessary, per the intended use section within the Package Inserts in the Kits. For any questions contact the firm's Global Support Service at: 1.877.785.2323 (US) 1.512.381.4397 (OUS/CANADA) 512.219.5114 [email protected] www.luminexcorp.com
Recall details
- Recall number
- Z-2775-2020
- Status
- Terminated
- Initiated by the company
- Jun 15, 2020
- Recalling firm
- Luminex Corporation
- Quantity
- Total = 3,016 systems (2240 Verigene Processor SP and 776 Verigene Processor SP Refurbished)
- Product code
- PAM
- FDA submissions
- K122514
- Distribution
- Worldwide distribution - US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CN, CT, DE, DC, FL,GA, ID, IN, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, MT, NB, NV, NE, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, RI, SC, SD, TN TX, UT, VT, VA, WA, WV, WY, and WI. The countries of Austria, Belgium, Bulgaria, Canada, Chile, China, France, Germany, Greece, Hong Kong, Italy, Japan, Republic of Korea, Kuwait, Macau, Mexico, Netherlands, New Zealand, Republic of Russia, Saudi Arabia, Singapore, Spain, Thailand, Turkey, United Arab Emirate, and United Kingdom.
Product description
Verigene Processor SP with assay: Luminex, VERIGENE Gram-Positive Blood Culture Nucleic Acid Test,BC-GP, IVD - Product Usage: utilizes a single-use disposable test consumables and a self-contained Verigene Test Cartridge for each sample tested.
Code information (models, lots, serial numbers)
Show the full code information (28k characters)
Verigene Processor SP, 10-0000-07, GTIN 00840487101483, Serial Numbers: 902, 907, 908, 909, 911, 912, 9098001, 9098002, 9098003, 9098005, 9098006, 9100009, 9100011, 9100012, 9100014, 9121015, 9121017, 9121018, 9121019, 9121021, 9121022, 9135024, 9135026, 9135027, 9135028, 9135029, 9163034, 9163035, 9163038, 9163039, 9163040, 9163041, 9163042, 9163043, 9163044, 9163045, 9163046, 9202047, 9202050, 9202051, 9202052, 9202053, 9202054, 9210055, 9210056, 9210057, 9210058, 9210059, 9210061, 9210062, 9230063, 9230064, 9230065, 9230066, 9230067, 9230068, 9230069, 9230070, 9230074, 9230075, 9255071, 9255072, 9255077, 9255079, 9255080, 9270080, 9270082, 9270084, 9270085, 9270086, 9270087, 9277088, 9277089, 9277090, 9277091, 9277092, 9277093, 9277094, 9288095, 9288096, 9288097, 9288102, 9293103, 9293106, 9293109, 9293111, 9298112, 9298116, 9298117, 9298118, 9313119, 9313120, 9313121, 9313122, 9313123, 9313124, 9320127, 9325138, 9325151, 9325152, 9336144, 9336145, 9336146, 9336147, 9336148, 9336149, 9336150, 9336155, 9336156, 9336157, 9336158, 9350159, 9350160, 9350161, 9350163, 9350164, 9350165, 10014001, 10014003, 10014012, 10035017, 10035105, 10035106, 10043023, 10043024, 10043025, 10043027, 10099029, 10099032, 10099033, 10099037, 10102046, 10110047, 10110048, 10110050, 10110051, 10110052, 10110053, 10110055, 10110056, 10110057, 10110058, 10110060, 10110061, 10110062, 10119063, 10119064, 10119065, 10119066, 10139080, 10139081, 10139082, 10139083, 10139084, 10162085, 10162092, 10162093, 10162094, 10182095, 10182096, 10182100, 10182101, 10182102, 10182103, 10182104, 10210105, 10210106, 10210107, 10210108, 10210109, 10210110, 10210111, 10210115, 10250117, 10250118, 10250119, 10250120, 10250121, 10250122, 10250123, 10276131, 10278128, 10278131, 10278132, 10278133, 10278163, 10295135, 10295136, 10295137, 10295138, 10295139, 10295140, 10295141, 10295142, 10295143, 10295144, 10295145, 10295146, 10295147, 10295149, 11021003, 11021008, 11101027, 11101028, 11122033, 11122034, 11152041, 11152042, 11152043, 11152044, 11152045, 11152046, 11152047, 11152048, 11152049, 11152050, 11152051, 11152052, 11152053, 11186054, 11186055, 11186056, 11186061, 11186062, 11186063, 11186064, 11186066, 11213071, 11213072, 11213073, 11213074, 11213075, 11213076, 11213078, 11213079, 11213080, 11213084, 11213085, 11236087, 11236089, 11236090, 11236091, 11236092, 11236093, 11250094, 11250095, 11250096, 11250098, 11250099, 11250100, 11250101, 11250103, 11250105, 11250106, 11250108, 11250109, 11276111, 11276112, 11276113, 11276114, 11276121, 11276122, 11276123, 11276124, 11276125, 11276127, 11276128, 11276129, 11306130, 11306131, 11306132, 11306133, 11306135, 11306136, 11306137, 11306141, 11306142, 11306144, 11306145, 11306146, 11306147, 11306148, 11306149, 11339151, 11339152, 11339153, 11339156, 11339157, 11339158, 11339159, 11339162, 11339163, 11339164, 12016001, 12016002, 12016003, 12016004, 12038011, 12038013, 12038014, 12038015, 12038016, 12038018, 12045021, 12045022, 12045023, 12045026, 12045027, 12045028, 12045029, 12045030, 12053034, 12053035, 12053036, 12053038, 12053039, 12053040, 12061041, 12061042, 12061043, 12061045, 12061048, 12061050, 12069052, 12069055, 12069056, 12069057, 12069058, 12069059, 12069060, 12079061, 12079062, 12079063, 12079064, 12079065, 12079066, 12079068, 12079069, 12079070, 12086071, 12086072, 12086073, 12086075, 12086076, 12086077, 12086078, 12086079, 12086080, 12088081, 12088082, 12088083, 12088084, 12088085, 12088086, 12088087, 12088088, 12088089, 12107091, 12107092, 12107093, 12107094, 12107095, 12107097, 12107098, 12111108, 12111109, 12111110, 12111111, 12111112, 12122056, 12128118, 12128119, 12128122, 12128123, 12128124, 12136125, 12136126, 12136127, 12136128, 12136134, 12136135, 12136136, 12142137, 12142138, 12142141, 12142142, 12145143, 12145144, 12145145, 12145146, 12145147, 12145148, 12145149, 12145150, 12145151, 12145152, 12145153, 12145154, 12156155, 12156156, 12156158, 12156160, 12156161, 12156162, 12156163, 12156164, 12156165, 12156166, 12170168, 12170169, 12170170, 12170171, 12170172, 12170173, 12170174, 12170175, 12170176, 12170177, 12175181, 12175182, 12175183, 12175184, 12175191, 12175192, 12175193, 12198194, 12198195, 12205211, 12205213, 12205214, 12205215, 12205216, 12205217, 12205218, 12212219, 12212220, 12212221, 12212222, 12212223, 12212224, 12212225, 12212226, 12212227, 12212228, 12219231, 12219232, 12219233, 12219234, 12219235, 12219236, 12219237, 12219238, 12223239, 12223240, 12223242, 12223243, 12229251, 12229252, 12229256, 12229258, 12233259, 12233262, 12233263, 12233264, 12233266, 12233267, 12236269, 12236270, 12236271, 12236272, 12236273, 12236274, 12236275, 12236276, 12236277, 12240280, 12240281, 12240282, 12240283, 12240284, 12240285, 12240286, 12248290, 12248291, 12248292, 12248296, 12254304, 12254305, 12254306, 12254307, 12254308, 12261309, 12261310, 12261311, 12261312, 12261314, 12261316, 12261317, 12261318, 12268319, 12268320, 12268321, 12268322, 12278340, 12278346, 12285349, 12285350, 12285351, 12285352, 12285353, 12285354, 12285355, 12285357, 12292360, 12292361, 12292362, 12292364, 12292365, 12292366, 12303370, 12303371, 12303372, 12303373, 12303375, 12303376, 12303377, 12303378, 12305379, 12305380, 12305382, 12305385, 12305386, 12305387, 12305388, 12310391, 12310392, 12310394, 12310397, 12317399, 12317405, 12324409, 12324410, 12324413, 12324414, 12324415, 12324416, 12324417, 12324418, 12333419, 12333420, 12333422, 12333423, 12333424, 12333425, 12333426, 12340429, 12340430, 12340431, 12340432, 12340433, 12340434, 12340435, 12340436, 12340437, 13003004, 13003005, 13003006, 13003007, 13003010, 13003011, 13003012, 13003013, 13008008, 13008009, 13008014, 13008015, 13008020, 13018025, 13018026, 13018027, 13018028, 13018030, 13024031, 13024032, 13024033, 13024034, 13024035, 13024036, 13024037, 13024038, 13024041, 13037042, 13037043, 13037044, 13037045, 13037046, 13037047, 13037048, 13037049, 13037050, 13042051, 13042052, 13042053, 13042055, 13042056, 13042057, 13042058, 13051061, 13051062, 13051063, 13051065, 13051069, 13051070, 13060071, 13060072, 13060073, 13060074, 13060075, 13060076, 13060077, 13060078, 13060079, 13066082, 13066083, 13066086, 13066089, 13072094, 13072095, 13072096, 13072097, 13072098, 13072099, 13072100, 13079105, 13079106, 13079107, 13079108, 13079109, 13087112, 13087113, 13087117, 13087118, 13087119, 13087120, 13092121, 13092122, 13092123, 13092124, 13092125, 13092127, 13092128, 13099131, 13099132, 13099133, 13099134, 13099135, 13099136, 13105142, 13105143, 13105144, 13105145, 13105146, 13105147, 13105148, 13105149, 13112153, 13112154, 13112155, 13112156, 13112157, 13112158, 13121167, 13121168, 13121169, 13121170, 13127175, 13127176, 13127177, 13127178, 13127179, 13133181, 13133183, 13133184, 13133185, 13133186, 13140194, 13140195, 13140198, 13140199, 13140200, 13154201, 13154202, 13154203, 13154204, 13154205, 13154206, 13154207, 13154208, 13161212, 13161213, 13161218, 13161219, 13161220, 13170221, 13170222, 13170223, 13170228, 13170229, 13178236, 13178237, 13178238, 13178239, 13186241, 13186242, 13186243, 13186244, 13186249, 13186250, 13193251, 13193252, 13193253, 13193254, 13193255, 13193256, 13193257, 13203266, 13203267, 13203268, 13210271, 13210272, 13210273, 13210279, 13210280, 13224283, 13224284, 13224285, 13224286, 13224289, 13227291, 13227292, 13227293, 13227294, 13227295, 13227296, 13227297, 13227298, 13227299, 13227300, 13238302, 13238303, 13238308, 13238309, 13245316, 13245317, 13245319, 13245320, 13263321, 13263322, 13263323, 13263324, 13263325, 13263326, 13263327, 13263328, 13263329, 13263330, 13274331, 13274333, 13274338, 13274339, 13274340, 13301341, 13301342, 13301343, 13301345, 13301346, 13301348, 13301349, 13315352, 13315354, 13315355, 13315356, 13315357, 13315360, 14006002, 14006003, 14006004, 14006005, 14006006, 14006007, 14006008, 14006010, 14013011, 14013012, 14013013, 14013014, 14013015, 14013016, 14013017, 14013018, 14013019, 14013020, 14043021, 14043022, 14043025, 14043026, 14043027, 14052031, 14052032, 14052035, 14052037, 14052038, 14052039, 14052040, 14062043, 14062044, 14062045, 14062046, 14062048, 14062049, 14069051, 14069053, 14069054, 14069055, 14069056, 14069057, 14069058, 14069059, 14083061, 14083062, 14083070, 14090071, 14090072, 14090073, 14090074, 14090075, 14090076, 14090077, 14090078, 14090080, 14111082, 14111083, 14111085, 14111088, 14111089, 14111090, 14118091, 14118094, 14118095, 14118096, 14118097, 14118098, 14118099, 14118100, 14139101, 14139102, 14139104, 14139105, 14139106, 14139107, 14142111, 14142112, 14142117, 14142118, 14142119, 14142120, 14155125, 14155126, 14155127, 14155129, 14155130, 14177131, 14177132, 14177136, 14177137, 14177138, 14177139, 14177140, 14177141, 14182144, 14182145, 14182146, 14182147, 14182148, 14182149, 14191151, 14191152, 14191156, 14191159, 14191160, 14196161, 14196162, 14196163, 14196164, 14196165, 14196166, 14196167, 14196168, 14196169, 14196170, 14211171, 14211172, 14211173, 14211174, 14211175, 14211176, 14211178, 14211179, 14218183, 14218184, 14218185, 14218186, 14218187, 14218188, 14218189, 14218190, 14224191, 14224192, 14224194, 14224195, 14224196, 14224197, 14224198, 14224200, 14230201, 14230206, 14230207, 14230208, 14230209, 14230210, 14237211, 14237212, 14237213, 14237214, 14237216, 14237217, 14237218, 14237219, 14237220, 14244222, 14244223, 14244224, 14244225, 14244226, 14244227, 14244228, 14244229, 14251231, 14251232, 14251235, 14251236, 14265242, 14265243, 14265244, 14265247, 14265248, 14265249, 14265250, 14307251, 14307252, 14307253, 14307254, 14307255, 14307257, 14307258, 14307259, 14307260, 14322261, 14322263, 14322264, 14322265, 14322266, 14322268, 15015001, 15015004, 15015008, 15015009, 15015010, 15020011, 15020015, 15020016, 15020019, 15020020, 15033023, 15033026, 15033027, 15033028, 15033029, 15033030, 15040031, 15040032, 15040034, 15040035, 15040036, 15040037, 15040038, 15040039, 15040040, 15055041, 15055042, 15055043, 15055045, 15055047, 15055048, 15055049, 15055050, 15075052, 15075054, 15075055, 15075056, 15075057, 15075058, 15075059, 15096061, 15096062, 15096064, 15096065, 15096066, 15096067, 15096068, 15131073, 15131074, 15131075, 15131076, 15131077, 15131078, 15131079, 15131080, 15208082, 15208083, 15208084, 15208085, 15208086, 15208087, 15208088, 15208090, 15320091, 15320092, 15320093, 15320094, 15320095, 15320096, 15320097, 15320098, 15320099, 15320100, 15327101, 15327102, 15327103, 15327104, 15327105, 15327106, 15327107, 15327108, 15327109, 15327110, 16049003, 16049004, 16049005, 16049006, 16049008, 16049009, 16049010, 16077011, 16077012, 16077014, 16077015, 16077016, 16077017, 16077018, 16077019, 16077020, 16125021, 16125022, 16125023, 16125024, 16125026, 16125027, 16125028, 16125029, 16125030, 16214031, 16214032, 16214033, 16214034, 16214035, 16214037, 16214038, 16214039, 16214040, 16242042, 16242044, 16242045, 16242046, 16242047, 16242048, 16242049, 16242050, 16277051, 16277052, 16277054, 16277055, 16277056, 16277057, 16277058, 16277059, 16277060, 16299063, 16299064, 16299065, 16299066, 16299067, 16299068, 16299069, 16299070, 16319071, 16319072, 16319073, 16319074, 16319075, 16334076, 16334077, 16334078, 16334079, 16334080, 16347081, 16347082, 16347083, 16347085, 16355086, 16355087, 16355088, 16355089, 16355090, 17017001, 17017002, 17017003, 17017004, 17017005, 17017006, 17017007, 17017008, 17017009, 17017010, 17048011, 17048012, 17048013, 17048014, 17048015, 17048016, 17048017, 17048018, 17048019, 17048020, 17062021, 17062022, 17062023, 17062024, 17062025, 17062026, 17062027, 17062028, 17062029, 17062030, 17096032, 17096033, 17096034, 17096035, 17096036, 17096037, 17096038, 17096039, 17096040, 17117041, 17117042, 17117043, 17117044, 17117045, 17117046, 17117047, 17117048, 17117049, 17117050, 17130051, 17130052, 17130053, 17130054, 17130055, 17130056, 17130057, 17130058, 17130060, 17157061, 17157062, 17157063, 17157064, 17157065, 17157066, 17157067, 17157068, 17157069, 17157070, 17177003, 17177004, 17177005, 17177006, 17177007, 17177008, 17177009, 17177010, 17188001, 17188002, 17188003, 17188004, 17188005, 17188006, 17188007, 17188008, 17188009, 17188010, 17215002, 17215003, 17215004, 17215005, 17215006, 17215007, 17234001, 17234002, 17234003, 17234004, 17234005, 17234006, 17234007, 17234008, 17234009, 17234010, 17242001, 17242002, 17242003, 17242004, 17242005, 17242007, 17242008, 17242009, 17242010, 17250001, 17250002, 17250003, 17250004, 17250005, 17250006, 17250007, 17250008, 17250009, 17250010, 17262001, 17262002, 17262003, 17262004, 17262005, 17262006, 17262007, 17262008, 17262009, 17278001, 17278002, 17278003, 17278004, 17278005, 17278006, 17278007, 17278008, 17278009, 17278010, 17284001, 17284002, 17284003, 17284004, 17284005, 17284007, 17284008, 17284009, 17284010, 17296001, 17296002, 17296003, 17296004, 17307004, 17307005, 17307006, 17307007, 17307008, 17307009, 17307010, 17333002, 17333003, 17333004, 17333005, 17340001, 17340002, 17340003, 17340004, 17340005, 17340006, 17340007, 17340008, 17347001, 17347002, 17347003, 17347004, 17347005, 17347006, 17347007, 17353001, 17353002, 17353003, 17353004, 17353005, 17353006, 17353007, 17353008, 17353009, 17353010, 17356001, 17356002, 17356003, 17356004, 17356005, 18022001, 18022002, 18022004, 18022005, 18023001, 18023002, 18023003, 18023004, 18023005, 18023006, 18023008, 18023009, 18023010, 18024001, 18024003, 18024004, 18024005, 18024006, 18024007, 18024008, 18024009, 18024010, 18026001, 18026002, 18026004, 18026005, 18026006, 18026007, 18026008, 18026009, 18026010, 18037002, 18037003, 18037004, 18037005, 18037006, 18037007, 18037008, 18037009, 18037010, 18039001, 18039002, 18039003, 18039004, 18039006, 18039008, 18039009, 18039010, 18046001, 18046002, 18046003, 18046004, 18046005, 18046006, 18046007, 18046008, 18046009, 18046010, 18054001, 18054002, 18054003, 18054004, 18054005, 18054006, 18054007, 18054008, 18054009, 18054010, 18057001, 18057002, 18057003, 18057004, 18057005, 18057006, 18057007, 18057008, 18057010, 18060001, 18060002, 18060003, 18060004, 18060005, 18060006, 18060007, 18060008, 18060009, 18072001, 18072002, 18072003, 18072005, 18072006, 18072007, 18072008, 18072010, 18079001, 18079002, 18079003, 18079004, 18079005, 18079006, 18079007, 18079008, 18079009, 18088001, 18088002, 18088003, 18088005, 18088006, 18088007, 18088008, 18088009, 18088010, 18099001, 18099002, 18099003, 18099004, 18099005, 18099007, 18099008, 18099009, 18099010, 18120001, 18120002, 18120003, 18120004, 18120005, 18120006, 18120007, 18120010, 18128001, 18128002, 18128003, 18128004, 18128005, 18128006, 18128007, 18128008, 18128009, 18128010, 18141001, 18141002, 18141003, 18141004, 18141005, 18141006, 18141008, 18141009, 18141010, 18150001, 18150002, 18150005, 18150007, 18150008, 18150009, 18150010, 18156001, 18156002, 18156003, 18156004, 18156005, 18156006, 18156007, 18156008, 18156009, 18156010, 18156012, 18156013, 18156014, 18156015, 18159001, 18159002, 18159003, 18159004, 18159005, 18159006, 18159007, 18159008, 18159009, 18159010, 18159011, 18159012, 18159013, 18172001, 18172002, 18172003, 18172004, 18172005, 18172006, 18172007, 18172008, 18172009, 18172010, 18172011, 18172012, 18172013, 18172014, 18172015, 18177001, 18177002, 18177003, 18177004, 18177005, 18177006, 18177007, 18177008, 18177009, 18177010, 18177011, 18177012, 18177013, 18177014, 18177015, 18191001, 18191002, 18191003, 18191004, 18191005, 18191007, 18191008, 18191009, 18191010, 18191011, 18191013, 18191014, 18191015, 18205001, 18205002, 18205003, 18205004, 18205005, 18205007, 18205008, 18205009, 18205010, 18205011, 18205012, 18205013, 18205014, 18205015, 18214001, 18214002, 18214003, 18214004, 18214005, 18214008, 18214009, 18214010, 18214011, 18214012, 18214013, 18214014, 18214015, 18226002, 18226003, 18226004, 18226005, 18226006, 18226008, 18226009, 18226010, 18226011, 18226013, 18226014, 18226015, 18235001, 18235002, 18235003, 18235004, 18235005, 18235006, 18235007, 18235008, 18235009, 18235010, 18235011, 18235012, 18235013, 18235014, 18235015, 18250001, 18250002, 18250003, 18250004, 18250005, 18250006, 18250007, 18250009, 18250010, 18250011, 18250012, 18250014, 18250015, 18255001, 18255002, 18255003, 18255004, 18255005, 18255007, 18255009, 18255012, 18255013, 18255014, 18264001, 18264003, 18264004, 18264005, 18264006, 18264008, 18264009, 18264010, 18264011, 18264012, 18264013, 18264014, 18264015, 18288001, 18288002, 18288003, 18288004, 18288005, 18288006, 18288007, 18288008, 18288009, 18288010, 18288011, 18288012, 18288013, 18288014, 18299001, 18299002, 18299004, 18299005, 18299006, 18299008, 18299009, 18299010, 18299011, 18299012, 18299013, 18299014, 18299015, 18318001, 18318002, 18318003, 18318004, 18318005, 18318006, 18318008, 18318009, 18318010, 18318011, 18318012, 18318013, 18318014, 18333001, 18333002, 18333003, 18333004, 18333005, 18333006, 18333007, 18333008, 18333010, 18333011, 18333012, 18333013, 18333014, 18344001, 18344002, 18344003, 18344004, 18344005, 18344006, 18344007, 18344009, 18344010, 18344011, 18344012, 18344013, 18344014, 18344015, 19023001, 19023002, 19023003, 19023004, 19023005, 19031001, 19031002, 19031003, 19031004, 19031005, 19031006, 19031007, 19031008, 19031009, 19031010, 19036001, 19036004, 19036005, 19036006, 19036007, 19036008, 19036009, 19036010, 19044001, 19044002, 19044003, 19044004, 19044005, 19044006, 19044007, 19044008, 19044009, 19044010, 19044011, 19044012, 19044013, 19044014, 19044015, 19057001, 19057003, 19057004, 19057005, 19057006, 19057007, 19057008, 19057009, 19057010, 19057011, 19057012, 19057014, 19057015, 19072002, 19072004, 19072005, 19072006, 19072007, 19072008, 19072009, 19072010, 19072011, 19072012, 19072013, 19072014, 19072015, 19084002, 19084003, 19084004, 19084005, 19084006, 19084007, 19084008, 19084009, 19084010, 19084014, 19092008, 19095001, 19095002, 19095003, 19095004, 19095005, 19095006, 19095007, 19095008, 19095009, 19095010, 19095011, 19095012, 19095013, 19095014, 19095015, 19123001, 19123003, 19123004, 19123006, 19123007, 19123008, 19123009, 19123010, 19140001, 19140002, 19140003, 19140004, 19140005, 19140006, 19140007, 19140008, 19140009, 19140010, 19155001, 19155002, 19155003, 19155004, 19155005, 19155006, 19155007, 19155008, 19155009, 19155010, 19192001, 19192002, 19192003, 19192004, 19192005, 19192006, 19192007, 19192008, 19192009, 19192010, 19206001, 19206002, 19206003, 19206004, 19206006, 19206007, 19206008, 19206009, 19206010, 19233001, 19233002, 19233003, 19233004, 19233005, 19233007, 19233008, 19233009, 19233010, 19260001, 19260002, 19260004, 19260005, 19260006, 19260007, 19260008, 19260009, 19260010, 19281001, 19281002, 19281003, 19281004, 19281005, 19282001, 19282002, 19282003, 19282004, 19282005, 19295001, 19295002, 19295003, 19295004, 19295005, 19295006, 19295007, 19295008, 19295009, 19295010, 19305001, 19305002, 19305003, 19305004, 19305005, 19305006, 19305007, 19305008, 19305009, 19305010, 19317001, 19317002, 19317003, 19317004, 19317005, 19317007, 19317008, 19317009, 19317010, 19325001, 19325002, 19325003, 19325004, 19325005, 19325006, 19325007, 19325008, 19325009, 19325010, 19343001, 19343002, 20007001, 20007002, 20007003, 20007004, 20007005, 20007006, 20007008, 20036001, 20036002, 20036003, 20036004, 20036005, 20036006, 20036007, 20036008, 20036009, 20036010, 20041001, 20048001, 20048002, 20048003, 20048004, 20048005, 20048006, 20048007, 20048008, 20048009, 20048010, 20058001, 20058002, 20058003, 20058004, 20058005, 20058006, 20058007, 20058008, 20058009, 20058010, 20071001, 20071002, 20071003, 20071004, 20071005, 20071006, 20071007, 20071008, 20071009, 20071010, 20086001, 20086002, 20086003, 20086004, 20086005, 20086006, 20086007, 20086008, 20086009, 20086010, 20094001, 20094003, 20094004, 20094005, 20094006, 20094007, 20094008, 20094009, 20094010, 20111001, 20111002, 20111003, 20111004, 20111005, 20111006, 20111007, 20111008, 20111009, 20111010, 20121002, 20121003, 20121004, 20121006, 20121007, 20121008, 20121010, 20133002, 20133004, 20133006, 20133007, D0903, D0908, D0909, D0912, D09135024, D09255073, D09313126, D09320128, D09320129, D09320130, D09320131, D09320132, D09320133, D09320134, D09320135, D09325135, D09325136, D09336157, D09336158, D908 Verigene Processor SP, Refurbished, 10-0000-07R, GTIN 00840487101490, Serial Numbers: 9100008, 9100013, 9121016, 9135023, 9135030, 9163032, 9163033, 9163036, 9202049, 9255076, 9270079, 9270081, 9270083, 9277087, 9288098, 9288099, 9288100, 9293104, 9293105, 9293107, 9293108, 9293110, 9298111, 9313125, 9325137, 9325139, 9325140, 9325141, 9325142, 9336143, 9336151, 9336152, 9336153, 9336154, 9350166, 10014002, 10014004, 10014005, 10014006, 10014007, 10014008, 10014009, 10014010, 10014011, 10035013, 10035014, 10035015, 10035016, 10035017, 10035018, 10035019, 10035020, 10043021, 10043022, 10043026, 10099030, 10099031, 10099035, 10099036, 10099038, 10099039, 10099040, 10099041, 10099042, 10099043, 10099044, 10102045, 10110059, 10119072, 10119075, 10119077, 10119078, 10139079, 10162086, 10162088, 10162090, 10162091, 10182097, 10182098, 10182099, 10210112, 10210113, 10210114, 10250115, 10250124, 10278125, 10278126, 10278127, 10278129, 10278130, 10278134, 10295150, 10295151, 10295152, 10295153, 10295154, 10336155, 10336156, 10336157, 10336158, 10336159, 10336160, 10336161, 10336162, 10336163, 10336164, 10350165, 10350166, 10350167, 10350168, 10350169, 10350170, 10350171, 10350172, 11021001, 11021002, 11021004, 11021005, 11021006, 11021007, 11021008, 11040009, 11040010, 11040011, 11040012, 11040014, 11040015, 11040016, 11068017, 11068018, 11068019, 11068020, 11068021, 11068022, 11068023, 11068024, 11101025, 11101026, 11101028, 11101029, 11101030, 11101031, 11101032, 11122035, 11122036, 11122037, 11122038, 11122039, 11122040, 11186057, 11186058, 11186059, 11186064, 11186065, 11186066, 11213067, 11213068, 11213069, 11213070, 11213082, 11213083, 11213085, 11213086, 11236087, 11236088, 11250090, 11250091, 11250092, 11250097, 11250104, 11276110, 11276115, 11276116, 11276117, 11276118, 11276119, 11276120, 11276126, 11306134, 11306136, 11306138, 11306139, 11306140, 11306143, 11306146, 11306147, 11339150, 11339155, 12016005, 12016006, 12016007, 12016008, 12016009, 12016010, 12038019, 12038020, 12045024, 12045025, 12053031, 12053032, 12053033, 12053035, 12053040, 12061044, 12061046, 12061047, 12061049, 12069051, 12069053, 12069054, 12069058, 12079067, 12086074, 12086077, 12088090, 12107096, 12107100, 12111101, 12111102, 12111103, 12111104, 12111105, 12111106, 12111107, 12128113, 12128114, 12128115, 12128116, 12128117, 12128120, 12128121, 12136130, 12136131, 12136132, 12136133, 12136135, 12142139, 12142140, 12145145, 12145146, 12145148, 12156161, 12156162, 12170167, 12170178, 12175179, 12175180, 12175185, 12175186, 12175187, 12175188, 12175189, 12175190, 12175192, 12175193, 12198196, 12198197, 12198198, 12198199, 12198200, 12198201, 12198202, 12198203, 12198204, 12198205, 12198206, 12198207, 12198208, 12205209, 12205210, 12205211, 12205212, 12205213, 12205214, 12219229, 12219230, 12223241, 12223244, 12223245, 12223246, 12223247, 12223248, 12229249, 12229250, 12229253, 12229254, 12229255, 12229256, 12229257, 12233260, 12233261, 12233265, 12240287, 12240288, 12248289, 12248293, 12248294, 12248295, 12248297, 12248298, 12254299, 12254300, 12254301, 12254302, 12254303, 12261313, 12261315, 12268323, 12268324, 12268325, 12268326, 12268327, 12268328, 12271329, 12271330, 12271331, 12271332, 12271333, 12271334, 12271335, 12271336, 12271337, 12271338, 12278339, 12278341, 12278342, 12278343, 12278344, 12278345, 12278347, 12278348, 12285356, 12285357, 12285358, 12292359, 12292363, 12292368, 12303369, 12303370, 12303371, 12303372, 12303373, 12303374, 12305381, 12305383, 12305384, 12310389, 12310390, 12310391, 12310392, 12310393, 12310396, 12310398, 12317400, 12317401, 12317402, 12317403, 12317404, 12317406, 12317407, 12317408, 12324411, 12324412, 12333426, 12333427, 12333428, 12340438, 13003003, 13003008, 13008011, 13008013, 13008014, 13008017, 13018024, 13018027, 13018028, 13018029, 13024039, 13024040, 13037041, 13037048, 13042059, 13042060, 13051062, 13051063, 13051064, 13051065, 13051066, 13051067, 13051068, 13060080, 13066081, 13066084, 13066085, 13066088, 13066089, 13072091, 13072092, 13072093, 13079101, 13079102, 13079103, 13079104, 13079110, 13087111, 13087114, 13087115, 13092126, 13092129, 13092130, 13105150, 13112151, 13112152, 13112159, 13112160, 13121161, 13121162, 13121163, 13121164, 13121165, 13121166, 13127171, 13127172, 13127173, 13127174, 13127180, 13133182, 13133184, 13133187, 13133188, 13133189, 13133190, 13140191, 13140192, 13140193, 13140194, 13140195, 13140196, 13154209, 13154210, 13161211, 13161214, 13161215, 13161216, 13161217, 13170224, 13170225, 13170226, 13170227, 13170229, 13178231, 13178232, 13178233, 13178234, 13178240, 13186241, 13186245, 13186246, 13186247, 13186248, 13186249, 13193257, 13193258, 13193259, 13193260, 13203261, 13203262, 13203263, 13203264, 13203265, 13203269, 13203270, 13210274, 13210275, 13210276, 13210277, 13210278, 13224281, 13224282, 13224285, 13224287, 13224288, 13227296, 13238301, 13238304, 13238305, 13238306, 13238307, 13238309, 13238310, 13245311, 13245312, 13245313, 13245314, 13245315, 13245318, 13274332, 13274334, 13274335, 13274336, 13274337, 13274340, 13301341, 13301344, 13301347, 13301350, 13315351, 13315353, 13315354, 13315355, 13315358, 13315359, 14006001, 14006009, 14043022, 14043023, 14043024, 14043026, 14043028, 14043029, 14043030, 14052033, 14052034, 14052036, 14062041, 14062042, 14062047, 14062050, 14069051, 14069052, 14069053, 14069060, 14083063, 14083064, 14083065, 14083066, 14083067, 14083068, 14083069, 14090079, 14111081, 14111084, 14111086, 14111087, 14118092, 14118093, 14139103, 14139104, 14139106, 14139107, 14139108, 14139109, 14139110, 14142111, 14142112, 14142114, 14142115, 14142116, 14155121, 14155122, 14155123, 14155124, 14155128, 14177133, 14177134, 14177137, 14177138, 14182142, 14182143, 14182149, 14182150, 14191153, 14191154, 14191155, 14191157, 14191158, 14211177, 14211180, 14218181, 14218182, 14224199, 14230202, 14230203, 14230204, 14230205, 14237215, 14244221, 14244230, 14251233, 14251234, 14251237, 14251238, 14251239, 14251240, 14265241, 14265245, 14265246, 14307256, 14307257, 14307258, 14307259, 14307260, 14322261, 14322262, 14322267, 14322269, 15015004, 15020012, 15020013, 15020014, 15020015, 15020016, 15020017, 15033021, 15033022, 15033024, 15033025, 15040033, 15055044, 15055046, 15075051, 15075053, 15075056, 15075060, 15131071, 15131072, 15131079, 15208081, 15208087, 15208088, 15208089, 15327109, 16049001, 16049002, 16077013, 16125025, 16242041, 16242043, 16277053, 16299061, 16299062, 16299066, 17062026, 17096031, 17096032, 17096033, 17096035, 17096037, 17117044, 17130059, 17177002, 17215005, 17215008, 17215009, 17262010, 17284001, 17284006, 17284009, 17296005, 17296006, 17296007, 17296008, 17296009, 17296010, 17307001, 17307003, 17333001, 17340004, 17347001, 18022003, 18023007, 18024002, 18026003, 18037001, 18037002, 18037003, 18037004, 18037005, 18039004, 18039005, 18046002, 18046003, 18046004, 18046005, 18046010, 18057009, 18060009, 18060010, 18072004, 18072005, 18072009, 18088003, 18088004, 18088005, 18099006, 18120008, 18128004, 18141007, 18150003, 18150004, 18150006, 18156011, 18159014, 18159015, 18172008, 18177012, 18191012, 18205006, 18214007, 18226001, 18226007, 18226011, 18226012, 18226013, 18235005, 18235007, 18235009, 18250008, 18250013, 18255006, 18255008, 18255010, 18255015, 18264002, 18264006, 18264007, 18264008, 18288008, 18299007, 18299008, 18299012, 18318009, 18318015, 18333001, 18344004, 18344008, 19036002, 19036003, 19072006, 19072009, 19084011, 19084012, 19084013, 19084015, 19095001, 19095013, 19123005, 19295001, D09288101
Devices named in this recall
2 devices in the FDA device registry whose identifier appears in the recall's code information.
- Verigene® Processor SP 10-0000-07Luminex Corporation · Instrumentation for clinical multiplex test systems3 recalls15 reports
- Verigene® Processor SP (Refurbished) 10-0000-07RLuminex Corporation · Instrumentation for clinical multiplex test systems3 recalls6 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 85872).
- Z-2776-2020 · TerminatedVerigene Processor SP with assay Luminex, VERIGENE Gram-Negative Blood Culture Nucleic Acid Test, BC-GN, IVD - Product Usage: utilizes a single-use disposable test consumables and …
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page