Alaris Auto ID Module Model 8600, a modular infusion pump… recallZ-2723-2020 · CareFusion 303, Inc. · initiated Jun 30, 2020
Class I Terminated All recalls posted in August 2020
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from CareFusion 303, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Cleaning agents may damage Inter-Unit Interface (IUI) Connectors, which may lead to interruption of communication or power between PC Unit and modules, which may result in an infusion that stops with a CHANNEL DISCONNECTED message, and an audio/visual alarm on the PC Unit and an interruption of therapy or monitoring.
FDA root cause category: Device Design
What the company is doing
On 06/30/2020, Urgent Medical Device Recall Notifications were mailed to Directors of Biomedical Engineering, Nursing, Risk Management, Environmental Services. Actions for Clinical Users: 1.Inspect the IUI connectors on each PC unit and module prior to each use. DO NOT use a device with any damage, cracks, or surface contaminants (e.g., blue or green deposits, corrosion, cleaning residue deposition) on the IUI connectors, or with broken IUI connector pins. If damage, deposits, residue, or corrosion is noticed, take the device out of service and send the unit to Biomedical Engineering for repair. 2.Ensure back-up devices are readily available when infusing critical medications where interruptions could cause serious injury or death. Please also ensure that you have back-up monitoring devices (e.g., EtCO2, SpO2) that are critically important to the patient's care. 3.If the issue occurs, expedite a replacement PC unit and modules to restart the infusions and monitoring. 4. Complete and return the Customer Response Card. Actions for Device Cleaning Personnel: Firm will provide existing customers a System Cleaning Kit (including IUI connector covers and cleaning brushes), Service Bulletin 630, Best Practices for Cleaning, and an updated System User Manual Addendum in August 2020. Actions for Biomedical Engineering: Expedite inspection of the IUI connectors on each PC unit and module at your facility. In addition, develop a system to check the IUI connectors prior to returning a unit to use from storage or repair. Replace any IUI connector with damage, cracks, surface contaminants or contact or pin damage on the IUI connectors. System rental providers, please provide a copy of this letter and enclosures to your customers who are renting affected devices. Recall resources and customer support: www.bd.com/alaris-system-hardware-recall Additional questions contact Recall Support Center: 858-617-1316, [email protected]
Recall details
- Recall number
- Z-2723-2020
- Recall class
- Class I (most serious)
- Status
- Terminated
- Initiated by the company
- Jun 30, 2020
- Posted by the FDA
- Aug 6, 2020
- Recalling firm
- CareFusion 303, Inc.
- Quantity
- 17767
- Product code
- FRN
- FDA submissions
- K133532
- Distribution
- Worldwide distribution. US nationwide including Guam, Puerto Rico, American Samoa, and Military Pacific. Countries of: TR, QA, CA, AE, KW, MX, GI, IL, SA, CH, NL, BE, TW, ZA, PH, CN, AU, IN, FR, MY, NZ, GB, KR, PK, DE, SG, JP, CU, IT, and BH
Product description
Alaris Auto ID Module Model 8600, a modular infusion pump and monitoring system
Code information (models, lots, serial numbers)
During cleaning, updated cleaning instructions and IUI covers must be used for all serial numbers.
Other devices under the same FDA submission
4 devices cleared under K133532 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Alaris™ Pump Module, Model 8100Carefusion 303, Inc. · Pump, infusion1 recall2 deaths5,636 reports
- Alaris 8975Carefusion 303, Inc. · Infusion safety management software169 reports
- Alaris 9601Carefusion 303, Inc. · Pump, infusion1 recall0 reports
- Alaris™ Syringe Module, Model 8110 8100Carefusion 303, Inc. · Pump, infusion14 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 85691).
- Z-2716-2020 · TerminatedAlaris System PC Unit Model 8000, a modular infusion pump and monitoring system
- Z-2717-2020 · TerminatedAlaris System PC Unit Model 8015, a modular infusion pump and monitoring system
- Z-2718-2020 · TerminatedAlaris Pump Module Model 8100, a modular infusion pump and monitoring system
- Z-2719-2020 · TerminatedAlaris Syringe Module Model 8110, a modular infusion pump and monitoring system
- Z-2720-2020 · TerminatedAlaris PCA Module Model 8120, a modular infusion pump and monitoring system
- Z-2721-2020 · TerminatedAlaris EtCO2 Module Model 8300, a modular infusion pump and monitoring system
- Z-2722-2020 · TerminatedAlaris SpO2 Module Model 8210 and Model 8220, a modular infusion pump and monitoring system
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page