Skip to content

MEVION S250 Proton Radiation Treatment System recallZ-2686-2020 · Mevion Medical Systems, Inc. · initiated Jun 23, 2020

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Mevion Medical Systems, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The perforated screen that divides the inside of the process water tank has a very sharp edge across the top of the Heat Exchanger leading to potential injury for servicer.

FDA root cause category: Device Design

What the company is doing

This screen would not be accessed by normal user (only servicer). Firm is notifying servicing personnel and not consumer/user level. No customer letters/communication sent. Servicer will install edge guard on effected units. Technical Service Bulletin issued to servicer.

Recall details

Recall number
Z-2686-2020
Status
Terminated
Initiated by the company
Jun 23, 2020
Recalling firm
Mevion Medical Systems, Inc.
Quantity
8 units
Product code
LHO
FDA submissions
K120676
Distribution
MO, NJ, OK, FL, OH, DC, Netherlands

Product description

MEVION S250 Proton Radiation Treatment System; intended to deliver proton radiation treatment to patients

Code information (models, lots, serial numbers)

Lot #s: MEVION S250-0001 MEVION S250-0002 MEVION S250-0003 MEVION S250-0004 MEVION S250-0005 MEVION S250-0006 MEVION S250-0007 MEVION S250-0008  UDI: (01)00864366000100 (11) 131209 (01)00864366000100 (11)150526 (01)00864366000100 (11)150113 (01)00864366000100 (11)150430 (01)00864366000100 (11)160127 (01)00864366000100 (11)160603 (01)00864366000124 (11) 180220 (01)00864366000124 (11) 181219

Devices named in this recall

2 devices in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page