VITROS recallZ-2875-2020 · Ortho-Clinical Diagnostics · initiated Jul 13, 2020

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Ortho-Clinical Diagnostics or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Etrombopag Interference for VITROS Chemistry Products: TBIL Slides, VITROS BuBc Slides and VITROS XT TBIL-ALKP Slides

FDA root cause category: Other

What the company is doing

Ortho Clinical Diagnostics issued URGENT PRODUCT CORRECTION NOTIFICATION on 13 July 2020, customer letter (Ref. CL2018-181) via FedEx overnight courier and/ or ORTHO PLUS e-Communications and/ or US Postal Service Priority Mail (for PO Boxes only) to all customers who received one of the affected products in the previous 18 months. Foreign affiliates were informed by email on 13 July 2020 of the issue and instructed to notify their consignees of the issue and their required action until the updated the IFU is available. Letter states reason for recall, health risk and action to take: Instruction for Use will be update at a later date to indicate Eltrombopag as a known interferent. REQUIRED ACTIONS " Follow your normal laboratory procedures as you would for other known interferences." Complete the enclosed Confirmation of Receipt form no later than July 21, 2020. " Please forward this notification if the product was distributed outside of your facility. " Save this notification with your user documentation Questions, please contact Ortho Care" Technical Solutions Center at 1-800-421-3311

Recall details

Recall number
Z-2875-2020
Status
Terminated
Initiated by the company
Jul 13, 2020
Recalling firm
Ortho-Clinical Diagnostics
Quantity
US= 4076 OUS=949
Product code
CIG
FDA submissions
K190807
Distribution
Worldwide distribution - US Nationwide distribution and the countries of Australia, Belgium, Bermuda HM08, Brazil, Canada L3R 4G5, China, Denmark, France, Germany, Italy, Mexico, Norway, Poland, Portugal, Russia, Singapore 609917, Spain, Sweden, The Netherlands, United Kingdom.

Product description

VITROS XT Chemistry Products TBIL-ALKP Slides (ALKP is not affected) In Vitro Diagnostic use only - The TBIL test within the VITROS XT Chemistry Products TBIL-ALKP Slides quantitatively measures total bilirubin (TBIL) concentration in serum and plasma, Product Code: 6844296

Code information (models, lots, serial numbers)

All Current, Future and Past Lots  UDI: 10758750031801

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 86070).