LCP recallZ-2816-2020 · Synthes (USA) Products LLC · initiated Jul 21, 2020
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Synthes (USA) Products LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
May have one or more dimensional features out of tolerance includes: thread pitch diameter depth, thread minor diameter depth, plate thickness and end length out of tolerance that can potentially result in a reduced locking strength of an LCP screw
FDA root cause category: Process control
What the company is doing
DePuy Synthes issued Urgent Medical Device Removal letter dated July 21, 2020 stating reason for recall, health risk and action to take: Please take the following actions: 1. Immediately examine your inventory to determine if you have product subject to this action on hand and quarantine the subject product. 2. Complete the attached Verification Section (page 3 of this letter). Please include your facility name and address, account number, name of person completing the form, title, email address, telephone number and signature in the spaces provided. If you have product to be returned: " Call DePuy Synthes at 1-800-479-6329 to obtain a Return Authorization (RA) Number. " Return the completed Verification Section (page 3 of this letter) with the product to: Credit/Returns, DePuy Synthes, 1101 Synthes Avenue, Monument, CO 80132 " Send a copy of the completed Verification Section (page 3 of this letter) to DePuy Synthes by: Scan/email: [email protected] " Return any subject product within 10 business days. If you have no product to be returned: " Return the Verification Section (page 3 of this letter) to DePuy Synthes by: Scan/email: [email protected] 3. Forward this notice to anyone in your facility that needs to be informed. 4. If any of the subject product has been forwarded to another facility, contact that facility to arrange return. 5. Keep a copy of this notice visibly posted for awareness until the product subject to this action has been returned. While processing your returns, please maintain a copy of this notice with the product subject to this action and keep a copy for your records. Questions, please call 610-719-5450 or contact your DePuy Synthes Sales Consultant
Recall details
- Recall number
- Z-2816-2020
- Status
- Terminated
- Initiated by the company
- Jul 21, 2020
- Recalling firm
- Synthes (USA) Products LLC
- Quantity
- 191 units
- Product code
- HRS
- FDA submissions
- K011335
- Distribution
- US Nationwide distribution.
Product description
DePuy Synthes Trauma LCP One-Third Tubular Plate with Collar 8 Holes / 93mm - Product Usage: intended for fixation of fractures, osteotomies and non-unions of clavicle, scapula, olecranon, humerus, radius, ulna, pelvis, distal tibia and fibula, particularly in osteopenic bone. Part Number 241.381
Code information (models, lots, serial numbers)
Lot Number: 29P6036 UDI: 10886982166913
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- LCP 241.381Synthes GmbH · PLATE, FIXATION, BONE1 recall14 reports
Product lines
Other devices under the same FDA submission
10 devices cleared under K011335 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- LCP 241.351Synthes GmbH · PLATE, FIXATION, BONE15 reports
- LCP 241.371Synthes GmbH · PLATE, FIXATION, BONE12 reports
- LCP 241.401Synthes GmbH · PLATE, FIXATION, BONE12 reports
- LCP 241.361Synthes GmbH · PLATE, FIXATION, BONE11 reports
- LCP 241.421Synthes GmbH · PLATE, FIXATION, BONE4 reports
- LCP 441.381Synthes GmbH · PLATE, FIXATION, BONE3 reports
- LCP 441.351Synthes GmbH · PLATE, FIXATION, BONE2 reports
- LCP 241.331Synthes GmbH · PLATE, FIXATION, BONE1 report
- LCP 241.341Synthes GmbH · PLATE, FIXATION, BONE1 report
- LCP 441.371Synthes GmbH · PLATE, FIXATION, BONE1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.