Verigene Processor SP, Model Numbers 10-0000-07 and… recallZ-0476-2021 · Luminex Corporation · initiated Oct 13, 2020
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Luminex Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Luminex has determined that three (3) separate test fixtures used by field service engineers to service the VERIGENE SP Processors were procured from unapproved sources and, in some cases, were manufactured using unvalidated processes. These text fixtures were used to calibrate and verify VERIGENESP devices.
FDA root cause category: Nonconforming Material/Component
What the company is doing
On 10/13/2020, the firm sent the following notice: "As part of our corrective action plan, Luminex will be updating software on all VERIGENE SP instruments in the coming weeks, and systematically visiting every VERIGENE customer site to perform a supplemental preventative maintenance on your instruments. We will also likely be sending out a series of Customer Advisory Notifications with additional information on items related to our remediation efforts surrounding the Warning Letter, as well as updates related to other items identified for improved product performance. During this time, you may continue to use your VERIGENE products as directed in the package inserts for each product, or as otherwise directed through further Luminex communications. If you have any question, please do not hesitate to contact Luminex Global Support Services using the contact details below: Luminex Global Support Services 1-877-785-2323 (U.S. and Canada) +1-512-381-4397 (Outside U.S. and Canada) [email protected]"
Recall details
- Recall number
- Z-0476-2021
- Status
- Terminated
- Initiated by the company
- Oct 13, 2020
- Recalling firm
- Luminex Corporation
- Quantity
- 2838 units
- Product code
- PAM
- FDA submissions
- K143653
- Distribution
- Worldwide distribution. US nationwide and Kuwait, Austria, United Kingdom, Spain, France, Chile, Japan, Netherlands, Greece, Italy, Thailand, Germany, Turkey
Product description
Verigene Processor SP, Model Numbers 10-0000-07 and 10-0000-07R. Used with assay: Luminex, VERIGENE Gram-Positive Blood Culture Nucleic Acid Test,BC-GP, IVD - Product Usage: IVD sample processor, which utilizes a single-use disposable test consumable and a self-contained Verigene Test Cartridge for each sample tested.
Code information (models, lots, serial numbers)
Show the full code information (11k characters)
UDI 00857573006010 Model Number 10-0000-07, Serial Numbers: 10110051, 10110050, 10110048, 10110047, 10099033, 10099032, 10099029, 10043027, 10043025, 10043024, 10043023, 10035106, 10035105, 10014003, 10014001, 9350165, 9350164, 9350163, 9350161, 9350160, 9350159, 9336156, 9336150, 9336149, 9336148, 9336147, 9336146, 9336145, 9336144, 9325152, 9325151, 9325138, 9320127, 9313124, 9313123, 9313122, 9313121, 9313120, 9313119, 9298118, 9298117, 9298116, 9293111, 9293109, 9293106, 9293103, 9288102, 9288097, 9288096, 9288095, 9277094, 9277093, 9277092, 9277091, 9277090, 9277089, 9277088, 9270087, 9270086, 9270085, 9270084, 9270082, 9270080, 9255080, 9255079, 9255077, 9255072, 9255071, 9230075, 9230074, 9230070, 9230069, 9230068, 9230067, 9230066, 9230065, 9230064, 9230063, 9210062, 9210061, 9210059, 9210058, 9210057, 9210056, 9202052, 9202047, 9163046, 9163045, 9163044, 9163044, 9163043, 9163042, 9163041, 9163040, 9163039, 9163038, 9163035, 9163034, 9135029, 9135028, 9135027, 9135026, 9121022, 9121021, 9121019, 9121018, 9121017, 9121015, 911, 9100014, 9100012, 9100011, 9100009, 9098006, 9098005, 9098003, 9098002, 9098001, 907, 902, 10035017R, 10014012 Correct, 10014012, D09336158, 9336158, D09336157, 9336157, D09135024, 9135024, D0912, 912, D0909, 909, D908, D0908, 908, 18255014, 18150007, 18120010, 18120003, 17333003, 17250008, 17250005, 17250002, 17130060, 17017006, 16334077, 13154206, 13154205, 13154204, 13154203, 13154202, 13008020, 12333423, 12324418, 12324417, 12324416, 12324415, 12324414, 12254308, 12254307, 12254306, 12254305, 12248296, 12333422, 12333420, 12333419, 20212005, 20007006, 20007005, 20007004, 20007002, 20007001, 19317003, 19317002, 19295003, 19095010, 19031003, 18333008, 18318006, 18264010, 18250012, 18250011, 18205003, 18191003, 18060006, 18060006, 17250007, 17234003, 17234002, 17234001, 17017001, 16277051, 15015004, 14211179, 14211178, 14211174, 13178239, 13178238, 13178237, 13178236, 13154201, 13140200, 13140199, 12086072, 12079070, 12079069, 12079065, 9298112, 9210055, 9202054, 12136125, 12128124, 12045028, 12045027, 12045026, 11250109, 11186056, 11186055, 11152050, 11152049, 11152048, 10102046, 18046009, 12061043, 12061042, 18026001, 17250010, 17250009, 19295006, 13105147, 13024036, 13024035, 13024033, 12219234, 12079068, 12079066, 11101027, 11276129, 17242010, 17242009, 17242008, 12107095, 12107094, 12107093, 12107092, 12107091, 18288014, 18288011, 17017003, 14090077, 13105148, 19095002, 13154208, 13154207, 13105146, 12285351, 12285350, 12219238, 12219233, 12053038, 12053036, 11339156, 11213084, 12240286, 12240285, 12240284, 12240283, 12240282, 12240281, 12240280, 12045030, 12045029, 10110062, 10110061, 10110060, 19192009, 19095005, 18054010, 18054009, 18054008, 17117050, 17117049, 17117048, 15327106, 15327105, 15327104, 15327103, 15327102, 15327101, 15320100, 15320099, 15320098, 15320097, 15320096, 15320095, 15320094, 15320093, 14251231, 14244229, 14244228, 13066086, 12205215, 11276128, 11276127, 9202053, 18255004, 18250010, 18172008, 13161218, 13105149, 13105145, 12219236, 12079064, 12079061, 11276121, 11276111, 11152052, 11152051, 10250123, 10250120, 10139083, 10139081, 10139080, 10099037, 14237220, 14237218, 14237216, 15040039, 15040038, 13099133, 13099132, 13099131, 12156156, 12156155, 17234007, 14090075, 14090074, 14090073, 13161219, 11276114, 11276113, 11276112, 10250121, 15327110, 14090076, 13042052, 13024038, 11250095, 11250094, 10182100, 13210280, 13210279, 18288012, 17017002, 13024037, 13024034, 12285353, 12285352, 12285349, 12223239, 12219237, 12219235, 12086071, 11186054, 11152053, 11152045, 10250122, 10139082 Model Number 10-0000-07R, Serial Numbers: 11186058, 9298111, 18255010, 17215009, 17096033, 16277053, 16125025, 15131079, 14191153, 13178234, 20094002, 19295001, 19260003, 19233003, 19206005, 19123005, 19123002, 19095013, 19095001, 19084015, 19084013, 19084012, 19084011, 19084001, 19072009, 19072006, 19057013, 19036003, 19036002, 18344008, 18344004, 18333001, 18318015, 18318009.18299012, 18299008, 18299007, 18299003, 18288008, 18264008, 18264007, 18264006, 18264002, 18255015, 18255008, 18255006, 18250013, 18250008, 18235009, 18235007, 18235005, 18226013, 18226012, 18226011, 18226007, 18226001, 18214007, 18205006, 18191012, 18177012, 18172008, 18159015, 18159014, 18156011, 18150006, 18150004, 18150003, 18141007, 18128004, 18120008, 18099006, 18088005, 18088004, 18088003, 18072009, 18072005, 18072004, 18060010, 18060009, 18057009, 18046010, 18046005, 18046004, 18046003, 18046002, 18046001, 18039007, 18039005, 18039004, 18037005, 18037004, 18037003, 18037002, 18037001, 18026003, 18024002, 18023007, 18022003, 17347002, 17347001, 17340004, 17333001, 17307003, 17307001, 17296010, 17296009, 17296008, 17296007, 17296006, 17296005, 17284009, 17284006, 17284001, 17278001, 17262010, 17215008, 17215005, 17215001, 17177002, 17177001, 17130059, 17117044, 17096037, 17096035, 17096032, 17096031, 17062026, 16347084, 16299066, 16299062, 16299061, 16242043, 16242041, 16214036, 16077013, 16049002, 16049001, 15327109, 15208089, 15208088, 15208087, 15208081, 15131072, 15131071, 15096063, 15075060, 15075056, 15075053, 15075051, 15055046, 15055044, 15055042, 15040033, 15033025, 15033024, 15033022, 15033021, 15020018, 15020017, 15020016, 15020015, 15020014, 15020013, 15020012, 14322269, 14322267, 14322267, 14322262, 14322261, 14307260, 14307259, 14307258, 14307257, 14307256, 14265246, 14265245, 14265241, 14251240, 14251239, 14251238, 14251237, 14251234, 14251233, 14244230, 14244221, 14237215, 14230205, 14230204, 14230203, 14230202, 14224199, 14218181, 14211180, 14211177, 14211173, 14211172, 14191158, 14191157, 14191155, 14191154, 14182150, 14182149, 14182143, 14182142, 14177138, 14177137, 14177135, 14177134, 14155128, 14155124, 14155123, 14155122, 14155121, 14142116, 14142115, 14142114, 14142112, 14142111, 14139110, 14139109, 14139108, 14139107, 14139106, 14139104, 14139103, 14118093, 14118092, 14111087, 14111086, 14111084, 14111081, 14090079, 14083069, 14083068, 14083067, 14083066, 14083065, 14083064, 14083063, 14069060, 14069053, 14069052, 14069051, 14062050, 14062047, 14062042, 14062041, 14052036, 14052035, 14052034, 14052033, 14043030, 14043029, 14043028, 14043026, 14043024, 14043023, 14043022, 14006009, 14006001, 13315359, 13315358, 13315355, 13315354, 13315353, 13315351, 13301350, 13301347, 13301344, 13301341, 13274340, 13274337, 13274336, 13274335, 13274334, 13274332, 13245318, 13245315, 13245314, 13245313, 13245312, 13245311, 13238310, 13238309, 13238307, 13238306, 13238305, 13238304, 13238301, 13227296, 13224290, 13224288, 13224287, 13224285, 13224282, 13224281, 13210278, 13210277, 13210276, 13210275, 13210274, 13203270, 13203269, 13203265, 13203264, 13203263, 13203262, 13203261, 13193260, 13193258, 13193257, 13186249, 13186248, 13186247, 13186246, 13186245, 13186241, 13178240, 13178233, 13178232, 13178231, 13170229, 13170227, 13170226, 13170225, 13170224, 13161217, 13161216, 13161215, 13161214, 13161211, 13154210, 13154209, 13140197, 13140196, 13140195, 13140194, 13140193, 13140192, 13140191, 13133190, 13133189, 13133188, 13133187, 13133184, 13133182, 13127180, 13127174, 13127173, 13127172, 13127171, 13121166, 13121165, 13121164, 13121163, 13121162, 13121161, 13112160, 13112159, 13112152, 13112151, 13105150, 13099140, 13099139, 13099138, 13099137, 13092130, 13092129, 13092126, 13087116, 13087115, 13087114, 13087111, 13079110, 13079104, 13079103, 13079102, 13079101, 13072093, 13072092, 13072091, 13066089, 13066088, 13066085, 13066084, 13066081, 13060080, 13060074, 13051068, 13051067, 13051066, 13051064, 13051063, 13051062, 13042060, 13042059, 13037048, 13037041, 13024040, 13024039, 13018029, 13018028, 13018027, 13018024, 13008017, 13008014, 13008013, 13008011, 13003008, 13003003, 12340438, 12333428, 12333427, 12333426, 12324412, 12324411, 12317408, 12317407, 12317406, 12317404, 12317403, 12317400, 12310398, 12310396, 12310393, 12310392, 12310391, 12310390, 12310389, 12305384, 12305383, 12305381, 12303374, 12303373.12303372, 12303371, 12303370, 12303369, 12292368, 12292363, 12292359, 12285358, 12285357, 12285356, 12278348, 12278347, 12278345, 12278344, 12278343, 12278342, 12278341, 12278339, 12271338, 12271337, 12271336, 12271335, 12271334, 12271333, 12271332, 12271331, 12271330, 12271329, 12268328, 12268327, 12268326, 12268323, 12261315, 12261313, 12254303, 12254302, 12254301, 12254300, 12254299, 12248298, 12248297, 12248295, 12248294, 12248293, 12248289, 12240288, 12240287, 12240279, 12236278, 12233265, 12233261, 12233260, 12229257, 12229256, 12229255, 12229254, 12229253, 12229250, 12229249, 12223248, 12223247, 12223246, 12223245, 12223244, 12223241, 12219230, 12219229, 12205214, 12205213, 12205212, 12205211, 12205210, 12205209, 12198208, 12198207, 12198206, 12198205, 12198204, 12198203, 12198202, 12198201, 12198200, 12198199, 12198198, 12198197, 12198196, 12175193, 12175192, 12175188, 12175187, 12175186, 12175185, 12175180, 12170178, 12170174, 12170167, 12156162, 12156161, 12145148, 12145147, 12145146, 12145145, 12142140, 12142139, 12142138, 12136135, 12136133, 12136132, 12136131, 12136129, 12128121, 12128120, 12128117, 12128116, 12128115, 12128114, 12128113, 12111107, 12111106, 12111105, 12111104, 12111103, 12111102, 12111101, 12107100, 12107096, 12088090, 12086077, 12086074, 12069058, 12069051, 12061049, 12061047, 12061044, 12053040, 12053035, 12053033, 12053032, 12053031, 12045025, 12045024, 12038020, 12038019, 12016010, 12016009, 12016008, 12016007, 12016006, 12016005, 11339155, 11339154, 11339150, 11306147, 11306146, 11306143, 11306140, 11306139, 11306138, 11306136, 11306134, 11276120, 11276119, 11276118, 11276117, 11276116, 11276115, 11276110, 11250104, 11250097, 11250092, 11250091, 11236088, 11236087, 11213086, 11213085, 11213083, 11213082, 11213077, 11213070, 11213069, 11213068, 11213067, 11186066, 11186065, 11186064, 11186060, 11186059, 11186057, 11122040, 11122039, 11122038, 11122037, 11122036, 11122035, 11101032, 11101031, 11101030, 11101029, 11101028, 11068022, 11068020, 11068019, 11068018, 11068017, 11040016, 11040015, 11040014, 11040012, 11040011, 11040010, 11040009, 11021008, 11021007, 11021006, 11021005, 11021004, 11021002, 11021001, 10350172, 10350171, 10350170, 10350169, 10350168, 10350167, 10350166, 10350165, 10336164, 10336163, 10336162, 10336161, 10336160, 10336159, 10336158, 10336157, 10336156, 10336155, 10295154, 10295153, 10295152, 10295151, 10295150, 10278134, 10278130, 10278129, 10278127, 10278126, 10278125, 10250124, 10250115, 10210114, 10210113, 10182099, 10182098, 10182097, 10162091, 10162090, 10162088, 10162086, 10139079, 10119077, 10119075, 10119072, 10110059, 10102045, 10099044, 10099043, 10099042, 10099041, 10099040, 10099039, 10099038, 10099036, 10099035, 10099031, 10099030, 10043026, 10043022, 10043021, 10035020, 10035019, 10035018, 10035014, 10035013, 10014011, 10014010, 10014009, 10014008, 10014007, 10014006, 10014005, 10014002, 9350166, 9336154, 9336153, 9336152, 9336151, 9336143, 9325142, 9325141, 9325140, 9325139, 9325137, 9313125, 9293110, 9293108, 9293107, 9293105, 9293104, 9288100, 9288099, 9288098, 9277087, 9270083, 9270081, 9270079, 9255076, 9202049, 9163036, 9163033, 9163032, 9135030, 9135023, 9121016, 9100013, 9100008, 10035017, D09288101, 15015004, 12079067, 11101026, 11101025, 11068024, 11068023, 10035016, 10035015
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Verigene® Processor SP 10-0000-07Nanosphere, Inc. · Real Time Nucleic Acid Amplification System3 recalls0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 86124).
- Z-0477-2021 · TerminatedVerigene Processor SP, Model Numbers 10-0000-07 and 10-0000-07R. Used with assay Luminex, VERIGENE Gram-Negative Blood Culture Nucleic Acid Test, BC-GN, IVD - Product Usage: IVD s…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.