ALTERA recallZ-2859-2020 · Globus Medical, Inc. · initiated Jul 22, 2020
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Globus Medical, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Specific lots of ALTERA implants have internal components that may have been manufactured using stainless steel rather than the specified cobalt chromium molybdenum alloy.
FDA root cause category: Nonconforming Material/Component
What the company is doing
A customer letter dated July 22, 2020 was mailed to customers. In this letter customers are instructed to immediately examine their inventory and determine if they have any of the affected implants. Alternatively, they may return all ALTERA inventory and Globus will sort out the affected implants and replace inventory with conforming implants. Report any adverse effects believed to be associated with the product to Globus Medical immediately by filling out the attached Response Card.
Recall details
- Recall number
- Z-2859-2020
- Status
- Terminated
- Initiated by the company
- Jul 22, 2020
- Posted by the FDA
- Aug 22, 2020
- Recalling firm
- Globus Medical, Inc.
- Product code
- MAX
- FDA submissions
- K140411
- Distribution
- Worldwide Distribution
Product description
ALTERA Spacer ALTERA Spacer, 10x31, 9-13mm, 8
Code information (models, lots, serial numbers)
Part #: 1124.1112; Lot # / UDI #: BAX137PE / (01)00889095069686(10)BAX137PE; BAX167SE / (01)00889095069686(10)BAX167SE; BAX178FE / (01)00889095069686(10)BAX178FE; BAX208LE / (01)00889095069686(10)BAX208LE; BAX250CE / (01)00889095069686(10)BAX250CE; BAX348CE / (01)00889095069686(10)BAX348CE; BAX431NE / (01)00889095069686(10)BAX431NE
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- ALTERA 1124.1112Globus Medical, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall1 report
Product lines
Other devices under the same FDA submission
22 devices cleared under K140411 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- ALTERA 1124.1111Globus Medical, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall4 reports
- ALTERA 1124.1011Globus Medical, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall2 reports
- ALTERA 1124.1013Globus Medical, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall0 reports
- ALTERA 1124.1015Globus Medical, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall0 reports
- ALTERA 1124.1032Globus Medical, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall0 reports
- ALTERA 1124.1033Globus Medical, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall0 reports
- ALTERA 1124.1035Globus Medical, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall0 reports
- ALTERA 1124.1113Globus Medical, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall0 reports
- ALTERA 1124.1115Globus Medical, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall0 reports
- ALTERA 1124.1132Globus Medical, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall0 reports
- ALTERA 1124.1133Globus Medical, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall0 reports
- ALTERA 1124.1135Globus Medical, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 86164).
- Z-2850-2020 · TerminatedALTERA Spacer ALTERA Spacer, 10x31, 8-12mm, 8 (sterile)
- Z-2851-2020 · TerminatedALTERA Spacer ALTERA Spacer, 10x26, 8-12mm, 8
- Z-2852-2020 · TerminatedALTERA Spacer ALTERA Spacer, 10x26, 9-13mm, 8
- Z-2853-2020 · TerminatedALTERA Spacer ALTERA Spacer, 10x26, 10-14mm, 8
- Z-2854-2020 · TerminatedALTERA Spacer ALTERA Spacer, 10x26, 12-16mm, 8
- Z-2855-2020 · TerminatedALTERA Spacer ALTERA Spacer, 10x26, 9-13mm, 15
- Z-2856-2020 · TerminatedALTERA Spacer ALTERA Spacer, 10x26, 10-14mm, 15
- Z-2857-2020 · TerminatedALTERA Spacer ALTERA Spacer, 10x26, 12-16mm, 15
- Z-2858-2020 · TerminatedALTERA Spacer ALTERA Spacer, 10x31, 8-12mm, 8
- Z-2860-2020 · TerminatedALTERA Spacer ALTERA Spacer, 10x31, 10-14mm, 8
- Z-2861-2020 · TerminatedALTERA Spacer ALTERA Spacer, 10x31, 12-16mm, 8
- Z-2862-2020 · TerminatedALTERA Spacer ALTERA Spacer, 10x31, 9-13mm, 15
- Z-2863-2020 · TerminatedALTERA Spacer ALTERA Spacer, 10x31, 10-14mm, 15
- Z-2864-2020 · TerminatedALTERA Spacer ALTERA Spacer, 10x31, 12-16mm, 15
- Z-2865-2020 · TerminatedALTERA Spacer ALTERA Spacer, 10x36, 8-12mm, 8
- Z-2866-2020 · TerminatedALTERA Spacer ALTERA Spacer, 10x36, 9-13mm, 8
- Z-2867-2020 · TerminatedALTERA Spacer ALTERA Spacer, 10x36, 10-14mm, 8
- Z-2868-2020 · TerminatedALTERA Spacer ALTERA Spacer, 10x36, 12-16mm, 8
- Z-2869-2020 · TerminatedALTERA Spacer ALTERA Spacer, 10x36, 9-13mm, 15
- Z-2870-2020 · TerminatedALTERA Spacer ALTERA Spacer, 10x36, 10-14mm, 15
- Z-2871-2020 · TerminatedALTERA Spacer ALTERA Spacer, 10x36, 12-16mm, 15
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page