Skip to content

AM1054 Randox Ammonia (NH3), Enzymatic UV Method recallZ-2966-2020 · Randox Laboratories, Ltd. · initiated Jul 30, 2020

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Randox Laboratories, Ltd. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Randox Ammonia reagents are being recalled from the field due to a positive bias of up to 140mol/l being observed on patient samples.

FDA root cause category: Under Investigation by firm

What the company is doing

The firm, Randox Laboratories Ltd., sent an "Urgent Medical Device Correction" letter dated 30th July 2020 to it customers. The letter described the product, problem and actions to be taken. The customers were instructed to do the following: - Discontinue use of, quarantine and discard any stock of AM1015, batch 517541 and AM1054, batch 512103 immediately. - Review results generated with the affected batches in line with the clinical profile of the patient. - Discuss the contents of this notice with your Medical Director. -Review your reagent inventory of these products and assess your laboratory needs for reimbursement for discarded inventory. -Complete and return the response form 12187-QA to [email protected] within five working days. If you have any questions or concerns please contact Randox Technical Services. Local at Tel: +1 304 728 2890 Toll Free 866 4 RANDOX Email: [email protected] and PR at Tel: 1-787-701-7000 Email: [email protected].

Recall details

Recall number
Z-2966-2020
Status
Terminated
Initiated by the company
Jul 30, 2020
Recalling firm
Randox Laboratories, Ltd.
Quantity
143 kits in total
Product code
JIF
FDA submissions
K030873
Distribution
US Distribution to states of: VA, UT, WV and country of : Canada

Product description

AM1054 Randox Ammonia (NH3), Enzymatic UV Method For the quantitative in vitro determination of Ammonia in plasma

Code information (models, lots, serial numbers)

Catalogue Number: AM1054 GTIN: 05055273200263 Lot #: 512103 batch 512103

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other products in the same recall event

The FDA groups related products into one recall event (event 86238).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page