Skip to content

RSP Standard HUMERAL SOCKET INSERT, 44MM, HXe-Plus, Part… recallZ-3021-2020 · Encore Medical, LP · initiated Aug 7, 2020

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Encore Medical, LP or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Product may be mislabeled with the size etched onto products differing from the size information shown on patient sticker labels. Risks associated with this issue may include the Surgeon implanting the incorrect size; significant physical impairment, discomfort and/or disruption of treatment of underlying condition; and the patient experiencing dislocation resulting in revision surgery.

FDA root cause category: Other

What the company is doing

Letters were sent to the firm's Field/Sales Agents by email beginning on August 7, 2020. The letters included instructions to take the following actions by August 21, 2020: 1. Pass the notice on to all those within your organization or to any organization where the potentially affected product has been transferred, 2.Contact Customer Service at 1-800-456- 8696 to place a replacement order and receive a Return Materials Authorization (RMA) number, and to be provided with a shipping label to return the affected product. 3. Return all affected devices using the RMA number, and 4. Report any adverse events related to this issue to Customer Service. Customers with questions should contact Christine Chesnutt at 512-834-6207 or [email protected].

Recall details

Recall number
Z-3021-2020
Status
Open, Classified
Initiated by the company
Aug 7, 2020
Recalling firm
Encore Medical, LP
Quantity
15 units
Product code
PHX
FDA submissions
K141006
Distribution
Distributed in the U.S. to: VA, MN, IN, RI, PA, NY, WA, KS, OK, FL, GA and CA

Product description

RSP Standard HUMERAL SOCKET INSERT, 44MM, HXe-Plus, Part No. 509-00-044

Code information (models, lots, serial numbers)

Part 509-00-044, Lot 380P1009A , UDI: 00888912144568

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

3 devices cleared under K141006 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 86317).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page