Skip to content

PermaFlo Flowable Composite REF-UP 948 Shade: A2 - Product… recallZ-0053-2021 · Ultradent Products, Inc. · initiated Aug 31, 2020

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Ultradent Products, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Due to a potential manufacturing issue (cross contamination), composite is non-homogeneous.

FDA root cause category: Under Investigation by firm

What the company is doing

On August 31, 2020, the firm sent an URGENT FIRST NOTICE PRODUCT RECALL" notification to US customers via UPS (International customer via email on September 1, 2020) informing them of the potential product contamination that may affect the performance characteristics of the product to include, micro-hardness, bond strength, and wear and may pose an increased risk of restoration failure. Customers are instructed to discontinue use of the product immediately. Customers are also instructed to: -Complete the Return Response Form enclosed with this letter and Fax 801-553-4609 or email to [email protected]. -Using the supplied shipping labels, return to Ultradent the completed form and all unused Product; and -Maintain a ledger of the patients treated with the Product, including the patients name and applicable treatment information. The Recalling Firm will send out replacement kits and issue an $100 credit to customer's accounts For questions regarding replacement product, contact Customer Service at 1-888-230-1420. For any questions regarding the recall, contact Customer Concerns at 801-553-4139.

Recall details

Recall number
Z-0053-2021
Status
Terminated
Initiated by the company
Aug 31, 2020
Recalling firm
Ultradent Products, Inc.
Quantity
412 syringes (206 kits)
Product code
EBF
FDA submissions
K974413
Distribution
Worldwide distribution - U.S. Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, FL, GA, IA, IL, KS, IN, LA, MA, MD, ME, MI, MT, NC, NJ, NM, NY, NV, OH, OK, PA, SC, TN, TX, UT, VT, WA, and WI. The countries of Australia, Germany Japan and Mongolia.

Product description

PermaFlo Flowable Composite REF-UP 948 Shade: A2 - Product Usage: is a light cured, radiopaque, methacrylate based, flowable composite. PermaFlo contains H1m average particle size with narrow upper limit particle distribution.

Code information (models, lots, serial numbers)

Part Number - 948: PermaFlo A2 Refill Lot Number - BJR5W UPC Code - 8832505103918 UDI Code - 00883205103918

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page