Cultura recallZ-0454-2021 · Merit Medical Systems, Inc. · initiated Sep 16, 2020

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Merit Medical Systems, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The flocked tip of the 100mm swab may break off in the nasophyaryngeal cavity during sample collection. This could result in soft tissue injury of the nasophyaryngeal cavity or a delay in treatment.

FDA root cause category: Device Design

What the company is doing

The firm sent an recall notification on September 16, 2020 to all affected consignees. In addition to informing consignees about the correction, the firm requested consignees take the following action: 1. Please refer to the attached instructions and/or linked instructional video when collecting samples with nasopharyngeal swabs. 2. Ensure that applicable personnel within your organization are made aware of this notice. 3. If the product has been further distributed to other facilities, institutions, or manufacturers, please ensure this notice is immediately shared with them. 4. If you have any questions or concerns regarding this communication or the products addressed, please contact your Merit Sales Representative or Merit Customer Service at (801) 208-4381.

Recall details

Recall number
Z-0454-2021
Status
Terminated
Initiated by the company
Sep 16, 2020
Recalling firm
Merit Medical Systems, Inc.
Quantity
3,385,442 units
Product code
KXG
Distribution
Worldwide distribution. US Nationwide, Australia, Canada, Colombia, Ghana, Ireland, Jamaica, Kenya, Panama, Qatar, Trinidad and Tobago.

Product description

MeritMedical Cultura Flex Swab, Sterile Flocked Collection Device, Ref C100F, Sterile, for the following unit count:1 unit - (01)00884450491102, 25 unit - (01)10884450491109, 100 unit - (01)20884450491106

Code information (models, lots, serial numbers)

All Lots distributed prior to August 28, 2020.  1 unit - (01)00884450491102, 25 unit - (01)10884450491109, 100 unit - (01)20884450491106