Skip to content

KWIK-STIK recallZ-0303-2021 · Microbiologics Inc · initiated Oct 8, 2020

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Microbiologics Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

KWIK-STIK 6 Pack Klebsiella pneumoniae derived from NCTC 13443 (Catalog# 01145K) should contain the NDM-1 positive strain, but was confirmed to be NDM-1 negative.

FDA root cause category: Under Investigation by firm

What the company is doing

The recalling firm is sending consignees recall notification letters beginning 10/08/2020 via email. The correction communication was first extended to the first level of the distribution change (those who received product directly from Microbiologics). Distributors outside of the US will be contacted via email, and if follow up is necessary, the recalling firm's international sales team will work with each distributor to ensure this communication is extended to their end users. The recalling firm is working with its European Authorized Representative to coordinate this field safety corrective action in the EU.

Recall details

Recall number
Z-0303-2021
Status
Terminated
Initiated by the company
Oct 8, 2020
Recalling firm
Microbiologics Inc
Quantity
1 unit
Product code
JTR
FDA submissions
K861022
Distribution
Distribution in US, Brazil, Canada, Costa Rica, France, Germany, Ireland, Italy, Japan, New Zealand, Sweden, and United Kingdom.

Product description

Each KWIK-STIK unit contains a lyophilized microorganism pellet, an ampoule of hydrating fluid, and an inoculating swab. Each device is sealed within a laminated pouch that contains a desiccant to prevent adverse moisture accumulation. KWIK-STIK microorganisms are 3 passages or fewer from the reference culture and are guaranteed to recover when processed using the recommended media and incubation requirements. Available in packs of 2 or 6.

Code information (models, lots, serial numbers)

Catalog 01145K, Lot 1145-06-5  UDI: 30845357036811

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 86576).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page