PROCEED recallZ-0533-2021 · Ethicon, Inc. · initiated Oct 23, 2020
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Ethicon, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The firm is recalling the PROCEED Surgical Mesh device, because it is a sterile device, and a customer found a hair inside of the primary packaging of a single unit.
FDA root cause category: Under Investigation by firm
What the company is doing
Note: This recall is specific to this single lot of PROCEED Surgical Mesh. Health care practitioners who have treated patients using this product should follow those patients post-operatively in the usual manner with no additional action required. Product subject to the recall in your inventory can be identified by the product code information described in the above table. all unused PROCEED Surgical Mesh subject to this recall are required to be returned. Please utilize attachment 1 for assistance in identifying the product lots subject to this recall. ACTION REQUIRED: 1. Examine your inventory immediately to determine if you have product subject to this recall on hand and quarantine such product(s). 2. Remove the product subject to this recall and communicate the issue to relevant operating room or materials management personnel, or anyone else in your facility who needs to be informed. 3. If any product subject to this recall has been forwarded to another facility, contact that facility to arrange return. Please consider including a copy of this recall letter when communicating. 4. Complete the Business Reply Form (Attachment 2) confirming receipt of this notice and return to Ethicon sales representative or distributor by [INSERT DATE]. Please return the Business Reply Form even if you do not have products subject to this recall. 5. Follow instructions in the letter and immediately return unused impacted PROCEED Surgical Mesh subject to this recall to Ethicon sales representative or distributor. To receive credit reimbursement, customers must return product subject to this recall. Any non-affected product and any product returned after the date specified will not receive credit. 6. Keep this notice visibly posted for awareness until all product subject to this recall has been returned to Ethicon Korea. While processing your returns, please maintain a copy of this notice with the product subject to this recall and keep a copy for your re
Recall details
- Recall number
- Z-0533-2021
- Status
- Terminated
- Initiated by the company
- Oct 23, 2020
- Recalling firm
- Ethicon, Inc.
- Quantity
- 295 units distributed. Only 15 units are being recalled.
- Product code
- FTL
- FDA submissions
- K060713
- Distribution
- The impacted lot (295 units) was distributed to 27 countries worldwide excluding the US. However, the firm is only recalling the 15 units that were distributed in South Korea due to South Korea's Ministry of Food and Drug Safety (MFDS) requirements.
Product description
PROCEED Surgical Mesh Hernia Mesh Oval, 15 cm x 20 cm
Code information (models, lots, serial numbers)
Product Code - PCDG1 Product Lot - PHG118 GTIN Code - 10705031047716 Expiration Date: 07/31/2021
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- PROCEED PCDG1Johnson & Johnson International Inc. · Mesh, surgical, polymeric1 recall1 death637 reports
Product lines
Other devices under the same FDA submission
10 devices cleared under K060713 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- PROCEED PCDN1Johnson & Johnson International Inc. · Mesh, surgical, polymeric744 reports
- PROCEED PCDH1Johnson & Johnson International Inc. · Mesh, surgical, polymeric421 reports
- PROCEED PCDM1Johnson & Johnson International Inc. · Mesh, surgical, polymeric154 reports
- PROCEED PCDT1Johnson & Johnson International Inc. · Mesh, surgical, polymeric122 reports
- PROCEED PCDJ1Johnson & Johnson International Inc. · Mesh, surgical, polymeric103 reports
- PROCEED PCDR1Johnson & Johnson International Inc. · Mesh, surgical, polymeric80 reports
- PROCEED PCDB1Johnson & Johnson International Inc. · Mesh, surgical, polymeric54 reports
- PROCEED PCDW1Johnson & Johnson International Inc. · Mesh, surgical, polymeric43 reports
- PROCEED PCDD1Johnson & Johnson International Inc. · Mesh, surgical, polymeric42 reports
- PROCEED PCDL1Johnson & Johnson International Inc. · Mesh, surgical, polymeric30 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.