VITROS recallZ-0471-2021 · Ortho Clinical Diagnostics Inc · initiated Oct 16, 2020

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Ortho Clinical Diagnostics Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Product instability; biased ALKP values obtained from VITROS Chemistry Products Performance Verifier II Lot Q7692 when stored unopened at refrigerated temperatures

FDA root cause category: Under Investigation by firm

What the company is doing

Ortho Clinical Diagnostics issued letter IMPORTANT PRODUCT CORRECTION NOTIFICATION (CL2020-243) on 16 October 2020 via FedEx/US Priority Mail or ORTHO Plus e-Communications for delivery on October 19, 2020, stating reason for recall, health risk and action to take: If their laboratory used VITROS Performance Verifier II Lot Q7692 to verify performance of VITROS ALKP Slides and stores an inventory of unopened vials in the refrigerator: Discard any remaining refrigerated inventory of unopened VITROS Performance Verifier II Lot Q7692. If their laboratory does not store inventory of VITROS Performance Verifier II Lot Q7692 in the refrigerator or does not use Lot Q7692 to verify performance of VITROS ALKP Slides: No Action Is Required. To verify performance of VITROS ALKP Slides, use frozen (never refrigerated) inventory of VITROS Performance Verifier II Lot Q7692 or switch to another lot of VITROS Performance Verifier II. Ortho will replace or credit your account as indicated on your Confirmation of Receipt form Retain this notification as verification of the revised instructions. OUS notified by email on 09/14/20 to global affiliates for distribution to distributors and customers.

Recall details

Recall number
Z-0471-2021
Status
Open, Classified
Initiated by the company
Oct 16, 2020
Recalling firm
Ortho Clinical Diagnostics Inc
Quantity
10788 total: US 4680; OUS 6108
Product code
JJY
FDA submissions
K904768
Distribution
Worldwide distribution - US Nationwide distribution and the countries of Australia, Belgium, Bermuda, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico, Norway, Poland, Portugal, Russia, Singapore, Spain, Sweden, The Netherlands, United Kingdom.

Product description

VITROS Chemistry Products Performance Verifier II-For in vitro diagnostic use only.VITROS Chemistry Products Performance Verifier is an assayed control used to monitor performance on VITROS 250/350/5,1 FS/4600/XT 3400 Chemistry Systems and the VITROS 5600/XT 7600 Integrated System - Product Code: 8231474

Code information (models, lots, serial numbers)

Lot Q7692 , EXP: 01-NOV-2021 UDI: 10758750004577