Skip to content

MAGNESIUM (Mg) recallZ-0546-2021 · Randox Laboratories Ltd. · initiated Nov 13, 2020

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Randox Laboratories Ltd. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Failing to meet the linearity performance claims quoted on the kit inserts-The linearity is reduced by approximately -30% for Serum/ Plasma and -50% for Urine

FDA root cause category: Under Investigation by firm

What the company is doing

Randox issued Urgent Medical Device Correction letter via e-mail to US Customers on 13 November 2020 stating reason for recall, health risk and action to take: Discontinue use of and discard any of the above immediately. " Review your reagent inventory of these products and assess your laboratories needs for reimbursement for discarded inventory. " Review results generated with the affected batches in line with the clinical profile of the patient. " Discuss the contents of this notice with your Medical Director. " Complete and return the response form 12187-QA to technical. [email protected] within five working days. Questions or concerns, please contact Randox Technical Services.

Recall details

Recall number
Z-0546-2021
Status
Terminated
Initiated by the company
Nov 13, 2020
Recalling firm
Randox Laboratories Ltd.
Quantity
1 kit
Product code
JGJ
FDA submissions
K974606
Distribution
US Nationwide distribution in the states of GA, NJ.

Product description

Randox Mg R1 - Product Usage: For the quantitative in vitro determination of Magnesium in serum, plasma and urine Catalogue Number: MG3880

Code information (models, lots, serial numbers)

Lot Number: 539734, Exp Date: 28 March 2022  UDI: 05055273204629

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 86827).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page