The cobas z 480 analyzer including dedicated software for… recallZ-0846-2021 · Roche Molecular Systems, Inc. · initiated Nov 19, 2020
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Roche Molecular Systems, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Dirty Lens May Cause Invalid or False Positive Results
FDA root cause category: Device Design
What the company is doing
Roche issued Urgent Medical Device Correction on 11/19/20 via UPS Ground (receipt required). Letter states reason for recall, health risk and action to take: Inform any operators that use the cobas z 480 analyzer of the potential hazards associated with this issue and provide a copy of this notification, as appropriate. " Ensure the AD-plates are sealed properly before performing amplification/detection on your cobas z 480 analyzer(s). " Refer to the cobas 4800 System User Assistance/cobas 4800 System Operators Manual in the section To seal the microwell plate for detailed instructions on how to properly seal the amplification/detection plate prior to loading on the cobas z 480 analyzer. Note: Review the additional warning messages in the General Safety sections that highlight recommendations for avoiding incorrect handling, risk of incorrect results, or contamination of the analyzer due to lack of proper sealing of the microwell plate. " Follow laboratory standard operating procedures to investigate the potential for false positive results for assays where a change in result reporting could impact patient management. " Consult with the physician or pathologist at your facility to determine specific clinical implications for your patients Contact the Roche Support Network Customer Support Center at 1-800-526-1247 if you determine that you are observing false positive results, experiencing a higher than usual number of invalid results, or observing that an unsealed/improperly sealed plate was used during a run. " Complete the attached faxback form (TP-01092) and fax or email it according to the instructions on the form. " If your facility has distributed cobas z 480 analyzers to another site, please ensure this UMDC is provided to that site. " File this UMDC for future reference. A Roche Field Service Engineer (FSE) will perform an inspection of the Optical Lens Detection Unit on your cobas z 480 analyzer(s). If the optical lens is found to be dir
Recall details
- Recall number
- Z-0846-2021
- Status
- Terminated
- Initiated by the company
- Nov 19, 2020
- Recalling firm
- Roche Molecular Systems, Inc.
- Quantity
- 597 units US; 3411 units OUS
- Product code
- MAQ
- FDA submissions
- P100020
- Distribution
- Worldwide distribution - US Nationwide distribution and the countries of Canada, Germany, Italy, Spain, United Kingdom, France, Austria, Turkey, Belgium, Poland, Portugal, Greece, Netherlands, Switzerland, Czech Republic, Romania, Sweden, Norway, Denmark, Hungary, Ireland, Slovakia, Croatia, Bulgaria, Finland, Lithuania, Bosnia-Herz, Slovenia, Iceland, Estonia, Australia, Cambodia, China, Hong Kong, India, Indonesia, Japan, Malaysia, Mongolia, Myanmar, New Zealand, Pakistan, Philippines, Singapore, South Korea, Taiwan, Thailand, Vietnam, Argentina, Aruba, Bahamas, Barbados, Belize, Brazil, Chile, Colombia, Costa Rica, Cuba, Dominican Rep., Ecuador, El Salvador, Guatemala, Honduras, Jamaica, Mexico, Panama, Paraguay, Peru, Uruguay, Venezuela, Iran, Israel, Kazakhstan, Kosovo, Lebanon, Macedonia, Oman, Russia, Saudi Arabia, South Africa, Sudan, Ukraine, UAE, Serbia, Montenegro, Algeria, Azerbaijan, Botswana, Burkina-Faso, Ethiopia, French Guayana, Gabon, Ghana, Guadeloupe, Lesotho, Martinique, Morocco, Nicaragua, Senegal, Sri Lanka, Tanzania, Tunisia, Uzbekistan, Zambia.
Product description
The cobas z 480 analyzer including dedicated software for IVD use is a plate-based automated amplification and detection system for nucleic acids. The analyzer is intended to be used as an in-vitro diagnostic or screening system providing real time PCR amplification and detection for human samples within the system. The cobas x 480 instrument and cobas z 480 analyzer are to be used by laboratory professionals trained in laboratory techniques and by instruction on the use of the system and analyzer. In the U.S. the cobas z 480 analyzer is used for automated amplification and detection purposes for the following assays: cobas 4800 BRAF V600 Mutation Test cobas Cdiff Test for use on the cobas 4800 System cobas CT/NG v2.0 Test cobas EGFR Mutation Test v2 cobas Factor II and Factor V Test cobas 4800 HPV Test cobas HSV 1 and 2 Test for use on the cobas 4800 System cobas KRAS Mutation Test cobas MRSA/SA Test for use on the cobas 4800 System
Code information (models, lots, serial numbers)
All units UDI: 04015630929016
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- cobas z 480 05200881001Roche Molecular Systems, Inc. · DNA PROBE, NUCLEIC ACID AMPLIFICATION, CHLAMYDIA1 recall0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page