Skip to content

CareLink Personal Software Therapy Management Software Tool… recallZ-0843-2021 · Medtronic Inc. · initiated Nov 12, 2020

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Medtronic Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Incomplete basal rate information may be transmitted leading to a potential for inappropriate change in therapy. During data upload, radio frequency (RF) interference can occur at any point from surrounding devices or equipment that may affect the data transfer between the insulin pump and the CareLink Personal or CareLink System Software.

FDA root cause category: Device Design

What the company is doing

On 11/19/2020, the firm sent an "URGENT FIELD SAFETY NOTIFICATION" Letter via email, fax, mail and/or phone to its U.S. Customer, and on 12/07/2020, the firm initiated O.U.S Notification using approved regional methods. The Notification Letter informs customers that after the release of an updated uploader software version the following issues/scenarios may result: 1. Some reports may be missing all programmed basal rates in the device. This may result in a blank value on the Device Settings report of the 24-hour basal total. 2. Some reports may only display certain programmed basal rates and some basal rates in the device may be missing from the reports. This may result in a miscalculation on the Device Settings report of the 24-hour basal total. The following reports may be affected by these issues include: Device Settings, Daily Review/Daily Details, Weekly Review, and Sensor & Meter Overview. These reports have been utilized to adjust therapy settings or to transfer settings from an old pump to a new pump. If all of the following are true, it could lead to either over- or under-delivery of insulin which may result in low blood glucose (hypoglycemia, loss of consciousness), and/or high blood glucose (hyperglycemia): a) adjustments were made to therapy settings or if therapy settings were transferred from one device to another, and b) these actions were taken based on a report generated using CareLink" uploader version 3.1.5.000 that has missing or incomplete basal rates, and c) you are using Manual Mode The Recalling Firm is instructing customers to: 1. Install the new CareLink" uploader version 3.1.6.000 available within the CareLink" software to each computer where you upload devices. To confirm your current uploader version or to learn how to install the latest uploader, see the FAQ section at the end of this letter. 2. If you have had insulin settings adjusted that may have been made based on an impacted CareLink" report, review your settings

Recall details

Recall number
Z-0843-2021
Status
Open, Classified
Initiated by the company
Nov 12, 2020
Recalling firm
Medtronic Inc.
Quantity
46,461 affected systems
Product code
MDS
FDA submissions
K151236
Distribution
U.S. Nationwide O.U.S. Worldwide Refer to Associated Documents and uploaded Consignee Lists/Spreadsheets. -Australia New Zealand -Canada -EMEA -Greater China -India -Japan -Latin America -South Korea -Southeast Asia -US

Product description

CareLink Personal Software Therapy Management Software Tool for Diabetes CareLink System Therapy Management Software Tool for Diabetes Software Uploader versions: 3.1.5.000 3.1.2.000

Code information (models, lots, serial numbers)

CareLinkTM Personal (MMT-7333) using CareLink Uploader (ACC-7350) UDI (GTIN/UPN): 00763000424121  CareLinkTM system (MMT-7350) using CareLink Uploader (ACC-7350) UDI (GTIN/UPN): 00763000424138

Devices named in this recall

2 devices in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page