Cook Medical Kwart Retro-Inject Ureteral Stent Set 7.0Fr.… recallZ-0960-2021 · Cook Inc. · initiated Jan 4, 2021
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Cook Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Manufactured with incorrect material, which may result in the stent being stiffer and potentially susceptible to degradation could potentially lead to breakage/fragmentation.
FDA root cause category: Nonconforming Material/Component
What the company is doing
Cook issued Urgent Medical Device Recall Letter on Jan 4 , 2021 via email/mail. Letter states reason for recall, health risk and action to take: If affected products are currently indwelling in a patient, consider scheduling a follow-up with the patient to remove and/or replace the product, based on the status and preferences of each individual patient. 2.Examine inventory immediately to determine if you have affected product(s) and quarantine any affected product that remains unused. Immediately cease all distribution and use of this product. 3.Return the affected product(s) to Cook Medical with a copy of the Acknowledgement and Receipt Form to receive a product credit. Refer to the Acknowledgement and Receipt Form for return instructions. Note: Unaffected products that are returned will not be credited. 4.Please complete the Acknowledgement and Receipt Form within 5 business days of receiving this letter. Even if you do not have affected product(s) on hand, you must still complete the Acknowledgement and Receipt Form and return via fax (812.339.7316) or email([email protected]). 5.This notice must be shared with appropriate personnel, including down to the user level, with in your organization or with any organization where the potentially affected devices have been transferred. 6.Immediately report adverse events to Cook Medical Customer Relations by phone at800.457.4500 or 812.339.2235, Monday through Friday between 7:30am and 5:00pm (EasternTime), or by email to: [email protected].
Recall details
- Recall number
- Z-0960-2021
- Status
- Terminated
- Initiated by the company
- Jan 4, 2021
- Recalling firm
- Cook Inc.
- Quantity
- 42 units
- Product code
- FAD
- FDA submissions
- K162109
- Distribution
- FL, KY, MD, NS, NY, PA, TX Foreign: Canada, Switzerland
Product description
Cook Medical Kwart Retro-Inject Ureteral Stent Set 7.0Fr. 22-32cm -Used for retrograde injection during Extracorporeal Shock Wave Lithotripsy (E.S.W.L.) and leaving an indwelling ureteral stent Reference Part Number (RPN): 003700 GPN: G14837
Code information (models, lots, serial numbers)
Lot Numbers: 13024536 13089895 NS13089897 UDI: (01)00827002148376(17)230303(10)13024536 (01)00827002148376(17)230327(10)13089895 (01)00827002148376(17)230327(10)NS13089897
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Kwart G14837Cook Incorporated · STENT, URETERAL1 recall0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page