EliA dsDNA Well - Product Usage: intended for the In-Vitro… recallZ-1107-2021 · Phadia US Inc · initiated Dec 14, 2020

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Phadia US Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Decreased values for EliA ANA Positive Control when using the EliA dsDNA Well lot BFA37/0142 may potentially cause false positive results (increase in patient test results).

FDA root cause category: Under Investigation by firm

What the company is doing

Phadia US Inc. (Initial Importer/Distributor) issued notification to consignees via phone and FedEx overnight delivery Dec. 17-18 . Letter states reason for recall, health risk and action to take: Customers are asked to perform the following steps: 1.Scrap EliA dsDNA Well lot BFA3Y/0142 and order a replacement free of charge 2. Assess test results from EliA dsDNA Well lot BFA3Y/0142 and determine if retesting of samples is warranted in accordance with internal operating procedures. 3. Complete the Acknowledgement Form in the Medical Device Field Safety Notice and return the response to the contact person described. Any questions concerning this issue, contact Phadia US Technical Support at 1-800-346-4364, option #2.

Recall details

Recall number
Z-1107-2021
Status
Terminated
Initiated by the company
Dec 14, 2020
Posted by the FDA
Feb 11, 2021
Recalling firm
Phadia US Inc
Quantity
365 kits US
Product code
LSW
FDA submissions
K072393
Distribution
US Nationwide distribution.

Product description

EliA dsDNA Well - Product Usage: intended for the In-Vitro quantitative measurement of IgG antibodies directed to dsDNA in human serum and plasma as an aid in the clinical diagnosis of systemic lupus erythematosus (SLE). Material Number: 14-5500-01

Code information (models, lots, serial numbers)

Carrier Lot Number: BFA3Y; Kit Lot Number: 0142; (UDI): 07333066019406

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.