Why it was recalled
The 5F Dignity CT Port kits were packaged with the incorrect size port. The kits include a 6.6F port instead of the correct 5F port. The inability to connect the catheter lumen to the port stem because of the incorrect size may not be noted until well into the procedure, after the catheter lumen is placed in the vessel. An alternate port will need to be obtained prolonging the procedure. A prolonged insertion procedure increases any risk associated with a surgical procedure.
FDA root cause category: Employee error
What the company is doing
On 1/13/2021 and 02/10/2021, Medical Components issued an Urgent Medical Device Removal notification to customers via letter notifying customers the 5F Dignity CT Port kits were packaged with the incorrect size port. The kits include a 6.6F port instead of the correct 5F port. (updated 3/11/2021)
Recall details
- Recall number
- Z-1184-2021
- Status
- Open, Classified
- Initiated by the company
- Jan 13, 2021
- Posted by the FDA
- Mar 8, 2021
- Recalling firm
- Medical Components, Inc dba MedComp
- Quantity
- 96 pieces total (updated 3/11/2021 additional 72 pieces)
- Product code
- LJT
- FDA submissions
- K132177
- Distribution
- US Nationwide distribution in the states of CA, DC, GA, IL, IN, MA, MI, NJ, NM, NY, OH, PA, TX. Updated 3/11/2021
Product description
Power Injectable CT Port insertion kits - Product Usage: is an implantable access device designed to provide repeated access to the vascular system.
Code information (models, lots, serial numbers)
Catalog Number: MRDP50AXS, Lot Number: MPDX180, UDI Number: 884908149753 (updated 3/11/2021 added lot number: MPCY660)
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- DIGNITY® MRDP50AXSMedical Components, Inc. · PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR1 recall3 reports
Product lines
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page