Medtronic Custom Pack CP Kinderset 1/4-1/4 MIT Pixie 1/B… recallZ-1197-2021 · Medtronic Perfusion Systems · initiated Feb 4, 2021
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medtronic Perfusion Systems or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Products being recalled due to potentially elevated level of bacterial endotoxin. Patients exposed to elevated levels of bacterial endotoxins may develop an acute inflammatory response.
FDA root cause category: Device Design
What the company is doing
A written Medical Device Correction letter will be mailed via UPS 2-day delivery to consignees who have received affected product according to Medtronic records. Consignees will be asked to complete the enclosed Customer Confirmation Form and email to [email protected]. Consignees will also be asked to return all unused affected product in your inventory to Medtronic. Contact Medtronic Customer Service at800-854-3570 to initiate a product return.
Recall details
- Recall number
- Z-1197-2021
- Status
- Terminated
- Initiated by the company
- Feb 4, 2021
- Posted by the FDA
- Mar 17, 2021
- Recalling firm
- Medtronic Perfusion Systems
- Quantity
- 5 units
- Product code
- DTN
- FDA submissions
- K183511
- Distribution
- Worldwide distribution: US nationwide, Afghanistan, Argentina, Chile, China, Colombia, Egypt, Germany, Iraq, Italy, Japan, Korea, Kyrgyzstan, Latvia, Nicaragua, Paraguay, Poland, Qatar, Romania, Tanzania, United Arab Emirates, United Kingdom, Uzbekistan, and Viet Nam
Product description
Medtronic Custom Pack CP Kinderset 1/4-1/4 MIT Pixie 1/B (PERFUSION M490817B 1/B). The Affinity Pixie Cardiotomy/Venous Reservoir (CVR) with Balance Biosurface is a single-use device designed to collect and store blood during extracorporeal circulation. Cardiotomy blood is collected, filtered, and defoamed before mixing with the venous blood, which is filtered. The primary blood-contacting surfaces of the CVR are coated with Balance Biosurface to reduce platelet activation and adhesion and preserve platelet function. (Generic Name: Cardiopulmonary bypass oxygenator).
Code information (models, lots, serial numbers)
Description: PERFUSION M490817B 1/B. Model Number: M490817B. GTIN: 00643169096271. Lot Number: 220886950.
Other devices under the same FDA submission
2 devices cleared under K183511 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Affinity Pixie™ BBP241Medtronic, Inc. · RESERVOIR, BLOOD, CARDIOPULMONARY BYPASS1 recall2 reports
- Affinity Pixie™ BBP211Medtronic, Inc. · OXYGENATOR, CARDIOPULMONARY BYPASS12 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 87271).
- Z-1190-2021 · TerminatedMedtronic Affinity Pixie Hollow Fiber Oxygenator and Cardiotomy/Venous Reservoir with Balance Biosurface (OXYGENATOR BBP241 AFFINITY PIXIE WCVR OU). The Affinity Pixie Cardiotomy/…
- Z-1191-2021 · TerminatedMedtronic Affinity Pixie BP241 Hollow Fiber Oxygenator and Cardiotomy/Venous Reservoir with Balance Biosurface (OXYGENATOR BBP241 W/CVR NI BB STERILE). The Affinity Pixie Cardioto…
- Z-1192-2021 · TerminatedCustom Perfusion System (CUSTOM PACK HY10J00R6 ECC SMALL CAR). The Affinity Pixie Cardiotomy/Venous Reservoir (CVR) with Balance Biosurface is a single-use device designed to col…
- Z-1193-2021 · TerminatedMedtronic Custom Pack Infant Tubingset Pixie Flow 0,8->1,5-1,8 1/B (PERFUSION M483203D 1/B). The Affinity Pixie Cardiotomy/Venous Reservoir (CVR) with Balance Biosurface is a sin…
- Z-1194-2021 · TerminatedMedtronic Custom Pack EVNP Clinical with AP40 1/B (PERFUSION M448036B 1/B). The Affinity Pixie Cardiotomy/Venous Reservoir (CVR) with Balance Biosurface is a single-use device de…
- Z-1195-2021 · TerminatedMedtronic Custom Perfusion Pack (CUSTOM PACK HY11B40R1 1/4 RESPIRATO). The Affinity Pixie Cardiotomy/Venous Reservoir (CVR) with Balance Biosurface is a single-use device designed…
- Z-1196-2021 · TerminatedCustom Perfusion System (CUSTOM PACK BB10H89R4 ECLR). The Affinity Pixie Cardiotomy/Venous Reservoir (CVR) with Balance Biosurface is a single-use device designed to collect and s…
- Z-1198-2021 · TerminatedMedtronic Custom Pack CP 1/4-1/4 Pediatric Set 1/B (PERFUSION M484901D 1/B). The Affinity Pixie Cardiotomy/Venous Reservoir (CVR) with Balance Biosurface is a single-use device de…
- Z-1199-2021 · TerminatedCustom Perfusion System (CUSTOM PACK BB7W03R8 CCSM NEON). The Affinity Pixie Cardiotomy/Venous Reservoir (CVR) with Balance Biosurface is a single-use device designed to collect a…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page