Skip to content

Enzyme linked immunosorbent assay, for herpes simplex… recallZ-1274-2021 · Siemens Healthcare Diagnostics, Inc · initiated Feb 8, 2021

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Siemens Healthcare Diagnostics, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Potential false positive HSV results. Siemens Healthcare Diagnostics, Inc. initiated a correction for ADVIA Centaur4P Herpes-2 lgG (HSV2) Onboard Stability and Calibration Frequency (Interval) Failures on the ADVIA Centaur XP System and ADVIA Centaur XPT System.

FDA root cause category: Under Investigation by firm

What the company is doing

In February 2021, Siemens will issue an Urgent Medical Device Correction notice to customers via letter stating that Siemens Healthcare Diagnostics Inc. has confirmed that the ADVIA Centaur HSV2 assay does not meet the 60 day onboard stability or the 28 day calibration frequency (interval) as stated in the ADVIA Centaur XP/XPT Herpes-2 IgG (HSV2).

Recall details

Recall number
Z-1274-2021
Status
Terminated
Initiated by the company
Feb 8, 2021
Posted by the FDA
Mar 18, 2021
Recalling firm
Siemens Healthcare Diagnostics, Inc
Quantity
390 kits total (Domestic 366 kits, Foreign 24 kits)
Product code
MYF
FDA submissions
K181334
Distribution
Domestic: AL, CA, FL, IL, MD, NJ, NY, TN, TX. Foreign: Nepal, Chile, Georgia, Spain.

Product description

Enzyme linked immunosorbent assay, for herpes simplex virus, hsv-2

Code information (models, lots, serial numbers)

***Updated as of 3/25/2021*** Lot #: B33255, B33269, B33413. UDI #:  (01)00630414016726(10)B33255(17)20210625,  (01)00630414016726(10)B33269(17)20210629, (01)00630414016726(10)B33413(17)20210819. Updated as of 3/25/2021

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page