Amicus Exchange Kit - Product Usage: intended for use in… recallZ-1344-2021 · Fenwal Inc · initiated Feb 25, 2021
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Fenwal Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Some lots of Amicus MNC Apheresis and Amicus Exchange kits have leaking centrifuge packs during a procedure on the Amicus Separator using therapeutic protocols only. The defect is a blood leak at the boot, elbow, or in the channel on the separation chamber of the centrifuge pack.
FDA root cause category: Under Investigation by firm
What the company is doing
A Recall notification letter titled, "URGENT - Medical Device Recall Amicus MNC Apheresis/Exchange Disposable Kits" and form titled, Amicus Therapeutic Kits Customer Reply Form were sent to consignees on 02/25/2021 via Fed Ex Overnight Express. The consignee requested their customers to: 1) Discontinue use of the affected products immediately by checking their inventory and quarantining all affected product at their facility. They were also informed to contact Fresenius Kabi Customer Service at 800-333-6925, Prompt 1 for replacement kits to ensure additional procedures can continue. 2) Inform potential users of the product in their organization of this notification. 3) Based on the options listed below, complete the attached response form and return via the fax number, email address, or the enclosed Fed Ex envelope and return address. Option A: If there is no remaining inventory of the affected product complete the response form and return it to Fresenius Kabi. Option B: If the customer has affected product in inventory and wishes to return it: Contact Fresenius Kabi Customer Service at 800-333-6925, Prompt 1 and request to return the affected product. They will assist the customer with placing a replacement order at that time. Option C: If the customer has a small number of kits in inventory and are willing to destroy the product on Fresenius Kabis behalf: Contact Fresenius Kabi Customer Service at 800-333-6925, Prompt 1 to identify if the customer has product to be destroyed and require replacement. They will provide a certificate of destruction for completion. Once completed, the certificate should be returned to Fresenius Kabi and replacements will be sent. Customers may be in a position that no alternate Device is available and therefore choose to continue with the use of the affected Amicus MNC Apheresis kits or Amicus Exchange kits (Therapeutics kits). For such cases, the following actions should be put in place until replacement products a
Recall details
- Recall number
- Z-1344-2021
- Status
- Terminated
- Initiated by the company
- Feb 25, 2021
- Recalling firm
- Fenwal Inc
- Quantity
- 2466 units
- Product code
- LKN
- FDA submissions
- K180615
- Distribution
- US Nationwide distribution in the states of GA, TX, TN, MD, CA, KS, MN, MA, NM, NC, LA, CO, KY, NV, PA, UT, MI, NJ, IL, MD, OH, NY, FL.
Product description
Amicus Exchange Kit - Product Usage: intended for use in the collection of blood components and mononuclear cells.
Code information (models, lots, serial numbers)
Product Code X6R2339; UDI: 04086000100151; Batch Numbers FA20F08147 (Exp. 06/07/2022), FA20H12145 (Exp. 08/11/2022), FA20J27157 (Exp. 10/26/2022)
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- AMICUS Exchange Kit X6R2339Fresenius Kabi AG · Separator, Automated, Blood Cell And Plasma, Therapeutic1 recall0 reports
Other devices under the same FDA submission
1 device cleared under K180615 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- AMICUS Exchange Kit - Therapeutics X6R2349Fresenius Kabi AG · Separator, Automated, Blood Cell And Plasma, Therapeutic2 recalls0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 87407).
- Z-1343-2021 · TerminatedAmicus MNC Apheresis Kit - Double Needle - Product Usage: intended for use in the collection of blood components and mononuclear cells.
- Z-1345-2021 · TerminatedAmicus Exchange Kit Therapeutics - Product Usage: intended for use in the collection of blood components and mononuclear cells.
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page