ATTUNE recallZ-1314-2021 · DePuy Orthopaedics, Inc. · initiated Feb 22, 2021
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from DePuy Orthopaedics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Globally the company is removing 37 lots affected lots for MBT & ATTUNE (S+) Tibial Bases due to potential damaged packaging. These implant components are utilized in Total Knee Arthroplasty. The affected tibial bases are designed to be used in cemented applications.
FDA root cause category: Packaging change control
What the company is doing
On 02/25/2021, Depuy Synthes issued an Urgent Medical Device Recall notices to customers via email. DePuy (Ireland) UC is initiating a voluntary removal for the 12 lots for ATTUNE (S+) Tibial Bases due to potential damaged packaging. These implant components are utilized in Total Knee Arthroplasty. The affected tibial bases are designed to be used in cemented applications.
Recall details
- Recall number
- Z-1314-2021
- Status
- Terminated
- Initiated by the company
- Feb 22, 2021
- Recalling firm
- DePuy Orthopaedics, Inc.
- Quantity
- 102 units total (Domestic: 6 units, Foreign: 86 units)
- Product code
- JWH
- FDA submissions
- K170806
- Distribution
- Worldwide distribution - US Nationwide distribution including in the states of AL, AZ, CA, CO, CT, DC, DE, FL, GA, ID, IL, KS, KY, LA, MA, MD, ME, MI, MN, MO, MT, NC, NE, NH, NJ, NY, OH, OK, OR, PA, SC, TN, TX, VA, VT, WA, WI, WV and the countries of Brazil, Austria, Belgium, Denmark, Estonia, France, Germany, Great Britain, Ireland, Israel, Italy, Macedonia, Mauritius, Netherlands, Norway, Poland, Romania, Slovakia, Slovenia, South Africa, Sweden, Switzerland, United Arab Emirates, China, Hong Kong, Japan, New Zealand, Korea, Singapore, Thailand.
Product description
ATTUNE FIXED BEARING KNEE - Product Usage: composed of individually packaged femoral, tibial and patellar components designed to replace the natural articular surface of the knee joint. The femoral component is a metal implant without a porous coating. The tibial component may be an all polyethylene component or comprised of a metal tibial base without porous coating, and a polyethylene insert and locking components. The patella component may be of an all polyethylene design.
Code information (models, lots, serial numbers)
Lot#: 9558300, Product Code: 150670003, GTIN 10603295492054 Lot#: 9555958, Product Code: 150670004, GTIN 10603295492061 Lot#: 9554006, Product Code: 150670005, GTIN 10603295492078 Lot#: 9556397, Product Code: 150670005, GTIN 10603295492078 Lot#: 9558618, Product Code: 150670006, GTIN 10603295492085 Lot#: 9557992, Product Code: 150670007, GTIN 10603295492092 Lot#: 9558238, Product Code: 150670008, GTIN 10603295492108 Lot#: 9558239, Product Code: 150670008, GTIN 10603295492108 Lot#: 9555613, Product Code: 150670009, GTIN 10603295492115
Devices named in this recall
7 devices in the FDA device registry whose identifier appears in the recall's code information.
- ATTUNE 1506-70-005Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER1 recall112 reports
- ATTUNE 1506-70-004Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER1 recall106 reports
- ATTUNE 1506-70-007Depuy (Ireland) · Prosthesis, knee, patellofemorotibial, semi-constrained, cemented, polymer + additive/metal/polymer + additive1 recall2 deaths74 reports
- ATTUNE 1506-70-006Depuy (Ireland) · Prosthesis, knee, patellofemorotibial, semi-constrained, cemented, polymer + additive/metal/polymer + additive1 recall70 reports
- ATTUNE 1506-70-003Depuy (Ireland) · Prosthesis, knee, patellofemorotibial, semi-constrained, cemented, polymer + additive/metal/polymer + additive1 recall64 reports
- ATTUNE 1506-70-008Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER1 recall43 reports
- ATTUNE 1506-70-009Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER1 recall9 reports
Product lines
Other devices under the same FDA submission
3 devices cleared under K170806 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- ATTUNE 1506-70-010Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER7 reports
- ATTUNE 1506-70-002Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER3 reports
- ATTUNE 1506-70-001Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 87433).
- Z-1315-2021 · TerminatedThe LCS COMPLETE P.F.C.TM SIGMA RP Mobile Bearing Total Knee System - Product Usage: consists of three components intended to replace the femoral, tibial, and patellar articular s…
- Z-1316-2021 · TerminatedThe ATTUNE Rotating Platform (RP) Total Knee System - Product Usage: consists of multiple components intended to replace the femoral, tibial, and patellar articular surfaces of the…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page