Bio-Console recallZ-1313-2021 · Medtronic Perfusion Systems · initiated Mar 3, 2021
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medtronic Perfusion Systems or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
An electrical component within the system controller module of certain Bio-Console 560 devices may exhibit electrical failure causing the device to stop pumping. Issues observed include the pump stopping, the user interface going blank, and smoking and producing a burning odor during use.
FDA root cause category: Component change control
What the company is doing
In the United States (US), beginning 03-Mar-2021, a written Urgent Medical Device Correction letter was mailed via UPS 2-day delivery to consignees who have received affected product according to Medtronic records. Consignees will be asked to complete and return a confirmation certificate, confirming they have received the notification.
Recall details
- Recall number
- Z-1313-2021
- Status
- Open, Classified
- Initiated by the company
- Mar 3, 2021
- Recalling firm
- Medtronic Perfusion Systems
- Quantity
- 387 devices (93 US)
- Product code
- DWA
- FDA submissions
- K131964
- Distribution
- Worldwide distribution. US Nationwide including Puerto Rico, Brazil, Brunei Darussalam, Canada, China, Colombia, Croatia, Ecuador, Guatemala, India, Israel, Italy, Japan, Kazakhstan, Korea, Mexico, Morocco, New Zealand, Russian Federation, Spain, Taiwan, Thailand, United Kingdom, and Viet Nam
Product description
Medtronic Bio-Console 560 Extracorporeal Blood Pumping Console, Model Numbers: 560BC, 560BC1, 560BCS, 560BCS1, R560BCS1. Used in cardiopulmonary bypass (CPB) procedures.
Code information (models, lots, serial numbers)
Show the full code information (5k characters)
560BC: GTIN 00885074539386, Serial Numbers: 560B100094, 560B100212, 560B100244, 560B100257, 560B100274, 560B100283, 560B100418, 560B100439, 560B100482 560BC1: GTIN 00613994450326, Serial Numbers: 560B100901, 560B101368 560BCS: GTIN 00613994258236, Serial Numbers: 560B100369, 560B100376, 560B100472, 560B100495, 560B100496 560BCS1: GTIN: 00613994450463, Serial Numbers: 560B100191, 560B100690, 560B100783, 560B100806, 560B100820, 560B100871, 560B100896, 560B101062, 560B101187, 560B101318, 560B101344, 560B101470, 560B101499, 560B101504, 560B101519, 560B101593, 560B101526 GTIN 00613994526052, Serial Numbers: 560B102610, 560B102725 GTIN 00643169201033, Serial Numbers: 560B101838 GTIN 00643169201651, Serial Numbers: 560B101834 GTIN 00643169315822, Serial Numbers: 560B100127, 560B101694, 560B101700, 560B101805, 560B101976, 560B101996, 560B102080, 560B102118, 560B102185, 560B102194, 560B102204, 560B102210, 560B102215, 560B102265, 560B102336, 560B102426, 560B102493, 560B102518, 560B102547, 560B102559, 560B102598, 560B102761, 560B102762, 560B102763, 560B102764, 560B102765, 560B102766 GTIN 00763000135447, Serial Numbers: 560B102711, 560B102717, 560B102718, 560B102720, 560B102721, 560B102722, 560B102723, 560B102724, 560B102726, 560B102727, 560B102730, 560B102731, 560B102732, 560B102733, 560B102734, 560B102735, 560B102736, 560B102737, 560B102738, 560B102739, 560B102740, 560B102741, 560B102742, 560B102743, 560B102745, 560B102746, 560B102747, 560B102748, 560B102750, 560B102751, 560B102752, 560B102754, 560B102755, 560B102756, 560B102757, 560B102758, 560B102760, 560B102767, 560B102768, 560B102769, 560B102770, 560B102771, 560B102772, 560B102773, 560B102774, 560B102775, 560B102776 GTIN 00763000159412, Serial Numbers: 560B102930, 560B102931, 560B102932, 560B102933, 560B102934 GTIN 00763000236526, Serial Numbers: 560B102778, 560B102779, 560B102780, 560B102781, 560B102782, 560B102783, 560B102784, 560B102785, 560B102786, 560B102787, 560B102788, 560B102789, 560B102790, 560B102792, 560B102793, 560B102794, 560B102795, 560B102796, 560B102797, 560B102798, 560B102799, 560B102800, 560B102801, 560B102802, 560B102803, 560B102804, 560B102805, 560B102806, 560B102807, 560B102809, 560B102810, 560B102811, 560B102812, 560B102813, 560B102815, 560B102817, 560B102818, 560B102819, 560B102820, 560B102821, 560B102822, 560B102823, 560B102824, 560B102825, 560B102826, 560B102827, 560B102828, 560B102829, 560B102830, 560B102831, 560B102832, 560B102833, 560B102834, 560B102835, 560B102836, 560B102837, 560B102838, 560B102839, 560B102840, 560B102841, 560B102842, 560B102843, 560B102844, 560B102845, 560B102846, 560B102847, 560B102848, 560B102849, 560B102850, 560B102851, 560B102852, 560B102853, 560B102854, 560B102855, 560B102856, 560B102857, 560B102858, 560B102859, 560B102860, 560B102861, 560B102862, 560B102863, 560B102864, 560B102865, 560B102866, 560B102867, 560B102868, 560B102869, 560B102870, 560B102871, 560B102872, 560B102873, 560B102874, 560B102875, 560B102876, 560B102877, 560B102878, 560B102879, 560B102880, 560B102882, 560B102883, 560B102884, 560B102885, 560B102886, 560B102887, 560B102888, 560B102889, 560B102890, 560B102891, 560B102892, 560B102893, 560B102894, 560B102895, 560B102896, 560B102897, 560B102898, 560B102899, 560B102900, 560B102901, 560B102902, 560B102903, 560B102907, 560B102908, 560B102909, 560B102910, 560B102912, 560B102913, 560B102914, 560B102915, 560B102916, 560B102917, 560B102918, 560B102919, 560B102920, 560B102921, 560B102922, 560B102923, 560B102924, 560B102925, 560B102926, 560B102927, 560B102928, 560B102929, 560B102946, 560B102947, 560B102948, 560B102949, 560B102950, 560B102951, 560B102952, 560B102953, 560B102954, 560B102955, 560B102956, 560B102957, 560B102958, 560B102959, 560B102960, 560B102961, 560B102962, 560B102963, 560B102964, 560B102965, 560B102966, 560B102967, 560B102968, 560B102969, 560B102970, 560B102971, 560B102972, 560B102973, 560B102974, 560B102975, 560B102976, 560B102977, 560B102978, 560B102979, 560B102980, 560B102981, 560B102982, 560B102983, 560B102984, 560B102985, 560B102986, 560B102987, 560B102988, 560B102989, 560B102990, 560B102991, 560B102992, 560B102993, 560B102994, 560B102995, 560B102996, 560B102997, 560B102998, 560B102999, 560B103000, 560B103001, 560B103002, 560B103003, 560B103004, 560B103005, 560B103006, 560B103007, 560B103008, 560B103009, 560B103010, 560B103011, 560B103012, 560B103013, 560B103014, 560B103015, 560B103016, 560B103017, 560B103018, 560B103019, 560B103020, 560B103021, 560B103022, 560B103023, 560B103024, 560B103025, 560B103026, 560B103027, 560B103028, 560B103029, 560B103031, 560B103032, 560B103033, 560B103034, 560B103035, 560B103036, 560B103037, 560B103038, 560B103039, 560B103040, 560B103041, 560B103042, 560B103043, 560B103044, 560B103045, 560B103046, 560B103047, 560B103048, 560B103049, 560B103050, 560B103051, 560B103052, 560B103053, 560B103054, 560B103055, 560B103057, 560B103058, 560B103059, 560B103060, 560B103061, 560B103062, 560B103069, 560B103070, 560B103071, 560B103072, 560B103073, 560B103074, 560B103075, 560B103077, 560B103078, 560B103079 R560BCS1, GTIN 00643169315822, Serial Numbers: 560B102009, 560B102355, 560B102367
Devices named in this recall
5 devices in the FDA device registry whose identifier appears in the recall's code information.
- NA 560BCS1Medtronic, Inc. · Control, pump speed, cardiopulmonary bypass1 recall135 reports
- Bio-Console® 560BCS1Medtronic, Inc. · Control, pump speed, cardiopulmonary bypass1 recall129 reports
- Bio-Console® 560BCS1Medtronic, Inc. · Control, pump speed, cardiopulmonary bypass1 recall1 death121 reports
- Bio-Console® 560BCS1Medtronic, Inc. · Control, pump speed, cardiopulmonary bypass1 recall30 reports
- Bio-Console® 560BC1Medtronic, Inc. · Control, pump speed, cardiopulmonary bypass1 recall21 reports
Product lines
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗