Skip to content

LIAISON 1-84 PTH Assay - Product Usage: is an in vitro… recallZ-1342-2021 · Diasorin Inc. · initiated Nov 23, 2020

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Diasorin Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Potential for system to produce falsely elevated parathyroid hormone (PTH) results from patient samples.

FDA root cause category: Under Investigation by firm

What the company is doing

A Recall notification letter titled, "Urgent DIASORIN CUSTOMER NOTIFICATION LETTER" was sent to consignees on 11/23/2020 via email. The letter instructs the consignee to stop using the LIAISON 1-84 PTH Assay (REF 310630) lot numbers 134909, 134909A, 134909B, and destroy any remaining inventory. They are asked to please complete the bottom portion of the letter and return to DiaSorin to obtain replacement product. The consignee is also instructed to forward the letter to anyone who may be impacted at their facility or anyone these kits have been distributed to outside of the consignee's organization.

Recall details

Recall number
Z-1342-2021
Status
Terminated
Initiated by the company
Nov 23, 2020
Recalling firm
Diasorin Inc.
Quantity
5,895 devices
Product code
CEW
FDA submissions
K150879
Distribution
Worldwide distribution - US Nationwide in the states of KY, OH, GA, PA, FL and the countries of Canada, Argentina, Bahrain, Bosnia & Herzegovnia, Iraq, Jordan, Morocco, Nepal, Russia, Serbia, Singapore, Taiwan, Thailand, Tunisia, Dubai, Italy, Israel, Brazil, Mexico, Australia, and Bulgaria.

Product description

LIAISON 1-84 PTH Assay - Product Usage: is an in vitro chemiluminescent immunoassay (CLIA) intended for the quantitative determination of parathyroid hormone (1-84) in human serum and EDTA plasma. Measurements of parathyroid hormone levels are used in the differential diagnosis of hypercalcemia and hypocalcemia resulting from disorders of calcium metabolism. The test has to be performed on the LIAISON Analyzer family.

Code information (models, lots, serial numbers)

REF 310630; lot numbers: 134909, 134909A and 134909B; UDI: 08056771600125

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page