Skip to content

Consist of a distal connection mechanism (either standard… recallZ-1367-2021 · Terumo Cardiovascular Systems Corporation · initiated Mar 2, 2021

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Terumo Cardiovascular Systems Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

There is a potential for the stainless steel locking plates on the device's sternal retractor locking mechanism to fracture.

FDA root cause category: Employee error

What the company is doing

A customer notification letter will be issued on 03/15/2021 by 2-day FedEx mailing to all (8 total) end users who have received affected product. Terumo CVS will provide a replacement device concurrent with the mailing. Customers will be advised to remove the affected product from use and to return product to Terumo CVS using the instructions provided in the customer letter.

Recall details

Recall number
Z-1367-2021
Status
Terminated
Initiated by the company
Mar 2, 2021
Recalling firm
Terumo Cardiovascular Systems Corporation
Quantity
12
Product code
MWS
Distribution
US Nationwide distribution in the states of NH, WI, NJ, PA, NV, MD, TN, PA.

Product description

Consist of a distal connection mechanism (either standard Terumo Ball Quick Connect or Terumo Shaft Quick Connect), an articulating arm with links, a handle for providing cable tension, and a mechanism to mount the device to a sternal retractor.

Code information (models, lots, serial numbers)

GTIN Code: (01)00699753450950 Catalog Number: T401163 Serial Numbers: T00393, T00394, T00395, T00399, T00403, T00405, T00406, T00407, T00408, T00410, T00411, T00412.

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page