The EOSedge system may acquire two simultaneous orthogonal… recallZ-1415-2021 · EOS Imaging · initiated Dec 22, 2020
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from EOS Imaging or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Inadequate images resizing and 2D measurement errors may occur when biplanar acquisition has been performed with patient orientation different from AP (Antero-Posterior).
FDA root cause category: Software design
What the company is doing
A recall notification letter titled, "Field Safety Notice (FSN1 REG-2020-0001-A) Safety information concerning the EOSedge imaging device, was sent to the US consignees on 12/22/2020 via certified mail, followed by a phone call to one of the consignees and hand delivery to the other U.S. consignee. The OUS consignees were provided notification beginning on 12/30/2020 either by sending of a certified mail plus an email to direct customer consignees or by sending of an email to distributors for countries where the EOSedge distribution is not made directly by EOS imaging, i.e. Australia and Spain. The consignee is requested to: 1) Please read the field safety notice. 2) Follow the instructions provided in this field safety notice. 3) Please transfer this notice to whomever needs to be informed within the consignee's organization or in any organization to which the device may have been transferred. The instructions provided in the notice goes as follows: "EOSedge users are requested not to combine acquisition of patients in orientation different than AP with reference plane modification at REVIEW step. Users of the sterEOS workstation are requested to systematically display reference planes and check they are properly positioned before performing modeling or measurement, sending to PACS and carrying out a diagnosis. Same verification can be performed on sterEOS with images previously acquired, and correction can be applied for retrospective measurements if necessary."
Recall details
- Recall number
- Z-1415-2021
- Status
- Terminated
- Initiated by the company
- Dec 22, 2020
- Recalling firm
- EOS Imaging
- Quantity
- 11 devices
- Product code
- KPR
- FDA submissions
- K202394
- Distribution
- Distributed nationwide to MN, OH and internationally to Australia, Canada, Germany, Spain, and France
Product description
The EOSedge system may acquire two simultaneous orthogonal planar images for a standing or seated patient at low dose. The scan length is defined by the user and may cover the whole body or a specific area (spine, lower limbs, etc.). EOSedge consists of a gantry and an acquisition workstation. The gantry contains two orthogonal acquisition units, each of which comprises an X-ray source (high frequency generator + X-ray tube + collimator) and an X-ray detection system (detector + electronics). The acquisition workstation and its software control the generators and detectors. It is also used to enter patient data, via RIS/HIS or manually, to define acquisition parameters, process images and record images in DICOM 3.0 format. The intended use/indications for use, technological characteristics, and principles of operation of the EOSedge system are described below.
Code information (models, lots, serial numbers)
Product Reference:03663999000108; Serial Numbers: 8.1912.1008, 8.2001.1010, 8.2001.1011, 8.2002.1012, 8.1907.1005, 8.1909.1006, 8.2005.1014, 8.1905.1004, 8.1911.1007, 8.1912.1009, and 8.2006.1016
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- EOSedge™ 2.1Eos Imaging · System, X-Ray, Stationary1 recall0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page