Skip to content

MAMBA Flex recallZ-1429-2021 · Boston Scientific Corporation · initiated Mar 5, 2021

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Boston Scientific Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The firm initiated a removal of certain batches of MAMBA 135 and MAMBA Flex 135 Microcatheters due to incorrect labeling and packaging.

FDA root cause category: Labeling False and Misleading

What the company is doing

A Recall notification letter titled, "Medical Device Removal - Immediate Action Required" was sent to the OUS consignees on 03/05/2021 and the U.S. consignees on 03/08/2021 via overnight mail delivery. Customers outside of the US were either contacted via letter delivery or email. The recalling firm requested if the consignees identify any product from the listed batch within their inventory, to please segregate the product immediately and return it to Boston Scientific in accordance with the enclosed instructions. The consignees will receive replacement product for all removed product returned to Boston Scientific. Distributors are also informed in the letter that the removal depth is to the hospital level and the removal notification should be forwarded to their customers. The field removal letter provides customers with the product segregation and return instructions and accounts are asked to complete a Reply Verification Tracking Form indicating they have received the field removal and followed the included instructions.

Recall details

Recall number
Z-1429-2021
Status
Terminated
Initiated by the company
Mar 5, 2021
Recalling firm
Boston Scientific Corporation
Quantity
52 devices
Product code
DQY
FDA submissions
K190401
Distribution
Distributed nationwide to AL, AR, AZ, CO, FL, GA, IA, IL, KY, LA, MI, MN, MO, MS, NC, ND, NV, NY, OK, OH, PR, SC, TN, VA, WA, WI and internationally to Belgium, France, Great Britain, Israel, Italy, Lithuania, Netherlands, Poland, Puerto Rico, Switzerland

Product description

MAMBA Flex 135 Microcatheter. Packaged with protective sleeve over catheter, sealed in a Tyvek pouch, and is then labeled and placed in carton with IFU. A closure strip, top and side labels are applied to the carton.

Code information (models, lots, serial numbers)

UPN: H7493928713540 US (Green); Batches: 26578463 and 26578470; GTIN: 8714729940289; and Expiration Date: 4-Jan-23

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 87539).