Why it was recalled
Products show heavy growth of C. albicans after 24 hours of incubation due to misformulation (expected reaction is partial inhibition after 18-48 hours incubation) and reduce the identification of Neisseria contributing to a delay in time to definitive diagnosis or a delay in proper treatment
FDA root cause category: Under Investigation by firm
What the company is doing
BD issued Urgent Medical Device Correction letter on 3/17/21 to customers and distributors via FedEx tracked letter. Letter states reason for recall, health risk and action to take: 1.Immediately inspect your inventory for the specific catalog and lot numbers listed above. Destroy all products subject to the recall following your institutions process for destruction. 2.Share this recall notification with all users within your facility network to ensure they are also aware of this recall. 3.Complete the attached Customer Response Form and return to the BD contact noted on the form whether or not you have any of the impacted material so that BD may acknowledge your receipt of this notification per FDA requirements and subsequently process your product replacement, if required. BD will provide a product replacement for all destroyed inventory. If you require further assistance, please contact: BD Contact US Contact Information Outside of US BD Customer/ Technical Support 800-638-8663 Monday Friday 7:00am and 7:00pm EST For customers outside the US, contact the local BD representative or distributor.
Recall details
- Recall number
- Z-1377-2021
- Status
- Terminated
- Initiated by the company
- Mar 17, 2021
- Recalling firm
- Becton Dickinson & Co.
- Quantity
- 15550 EA
- Product code
- JTY
- FDA submissions
- K770287
- Distribution
- Worldwide distribution - US Nationwide distribution and the countries of Canada, Korea.
Product description
BD BBL Martin-Lewis Agar, 100 Ea - Product Usage: used for the isolation of pathogenic Neisseria from specimens containing mixed flora of bacteria and fungi. Catalog Number: 221558
Code information (models, lots, serial numbers)
Lot Numbers: 0317559 Exp. Date 2021-MAR-11; 0345695 Exp. Date 2021-APR-06 UDI: 01)30382902215586 (10)0317559(17)031121 (30)0100 (01)30382902215586 (10)0345695(17)040621 (30)0100
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- BD BBL™ Martin-Lewis Agar 221558Becton, Dickinson And Company · CULTURE MEDIA, FOR ISOLATION OF PATHOGENIC NEISSERIA1 recall1 report
Product lines
Other devices under the same FDA submission
1 device cleared under K770287 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- BD BBL™ Martin-Lewis Agar 221557Becton, Dickinson And Company · CULTURE MEDIA, FOR ISOLATION OF PATHOGENIC NEISSERIA1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 87546).
- Z-1378-2021 · TerminatedBD BBL Martin-Lewis Agar 20 Ea - Product Usage: use for the isolation of pathogenic Neisseria from specimens containing mixed flora of bacteria and fungi. Catalog Number: 221557
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page