Skip to content

CHESAPEAKE Stabilization System recallZ-1410-2021 · K2M, Inc · initiated Mar 16, 2021

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from K2M, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Interference condition between the 10mm Drill Guide Inserter and the inner shaft that mates with the inserters. As a result, inner shafts do not mate flush with the collet inserters, and splaying of the collet. If implant falls into the spinal cord may cause a spinal cord injury or Incomplete seating of the inserter to the implant could result in misalignment of the screws with the inserter screw guide, potentially causing malposition of the screws into the adjacent vertebral bodies

FDA root cause category: Under Investigation by firm

What the company is doing

Stryker Spine issued Urgent Medical Device Recal letter that issued on 3/16. Letter states reason for recall, health risk and action to take: Our records indicate Stryker previously distributed catalog no. 3608-90033, Lot MNAJ to your location in a Stryker Loaner Set. The nonconforming instrument was subsequently identified and quarantined during routine processing of loaner sets by Stryker at its Leesburg facility. This communication is being made for notification purposes only. The only action requested of you is to complete the Business Reply Form found on the following page and return it to the email address referenced on the form. If you have any questions or concerns, please contact Stryker Spine Regulatory Compliance at 201-749-8090. On behalf of Stryker we thank you sincerely for your help and support in completing this action by the target date March 19, 2021.

Recall details

Recall number
Z-1410-2021
Status
Terminated
Initiated by the company
Mar 16, 2021
Recalling firm
K2M, Inc
Quantity
1 unit
Product code
HRS
FDA submissions
K111439
Distribution
PA

Product description

Chesapeake Cervical-Ti 10mm Drill Guide Inserter- a multi-screw construct providing stability to the anterior column (cervical spine) Catalog number: 3608-90033

Code information (models, lots, serial numbers)

Lot Number: MNAJ GTIN: 10888857082571

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Sources and updates