Skip to content

Grace Medical REF S-4500 *** spineX BONE DUST COLLECTOR ***… recallZ-1521-2021 · Oto Med Inc · initiated Dec 7, 2020

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Oto Med Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

During internal Age Testing, holes were found in the sterile Tyvek pouches of the devices.

FDA root cause category: Package design/selection

What the company is doing

On 12/07/2020, the firm sent an "URGENT: PRODUCT RECALL NOTICE" Letter via FedEx Overnight (U.S.) and email (O.U.S.) to customer informing them that there is a potential for a hole to develop in the packaging during handling/shipment. Customers are instructed to: 1) Check their inventory for any holes or damage to the pouches, and disposition as specified below: -My inventory has been checked and included none of the affected products. -My inventory has been checked and has no damaged pouches; The product is not being returned. -My inventory has been checked and the product listed is being returned. If returning products, please obtain a Returned Authorization (RA) number. The RA number may be obtained from Customer Service (866-472-2363). In addition, the Response Form must be completed and returned to Grace Medical Group at RECALL COORDINATOR, OTO-MED, INC. c/o GRACE MEDICAL GROUP, INC., 8500 Wolf Lake Drive, Suite 110, Memphis, TN 38133, or by fax to 901-386-0950 or email at [email protected].

Recall details

Recall number
Z-1521-2021
Status
Open, Classified
Initiated by the company
Dec 7, 2020
Recalling firm
Oto Med Inc
Quantity
415 devices
Product code
MXP
Distribution
Worldwide distribution - U.S. Nationwide distribution including in the states of MN, IL, PA, WI, GA, SD, FL, AR, MA, AZ, TX, MO, NC, CA, UT, DE, OR, IN, CO, AL, OH, NY, WA, DC, TN, NH, VT, VA, SC, MI, WI, MS, IA, NJ, MD, KS, CT, LA, ME, KY, HI, TN, RI, NV, ID and the countries of Argentina, Brazil, Colombia, Chile, Hong Kong, South Africa, United Arab Emirates, Taiwan, Canada, India.

Product description

Grace Medical REF S-4500 *** spineX BONE DUST COLLECTOR *** 2 stainless steel filters - Product Usage: intended to be inserted into a suction tube during the early stages of otologic surgery to collect bone particles for future use.

Code information (models, lots, serial numbers)

UDI Code: 00850000126077 Lot Numbers: 019120 G(2) 0712119 G(2) 0912118 G(2) 1012119 G(2)

Devices named in this recall