Grace Medical REF S-4500 *** spineX BONE DUST COLLECTOR ***… recallZ-1521-2021 · Oto Med Inc · initiated Dec 7, 2020
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Oto Med Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
During internal Age Testing, holes were found in the sterile Tyvek pouches of the devices.
FDA root cause category: Package design/selection
What the company is doing
On 12/07/2020, the firm sent an "URGENT: PRODUCT RECALL NOTICE" Letter via FedEx Overnight (U.S.) and email (O.U.S.) to customer informing them that there is a potential for a hole to develop in the packaging during handling/shipment. Customers are instructed to: 1) Check their inventory for any holes or damage to the pouches, and disposition as specified below: -My inventory has been checked and included none of the affected products. -My inventory has been checked and has no damaged pouches; The product is not being returned. -My inventory has been checked and the product listed is being returned. If returning products, please obtain a Returned Authorization (RA) number. The RA number may be obtained from Customer Service (866-472-2363). In addition, the Response Form must be completed and returned to Grace Medical Group at RECALL COORDINATOR, OTO-MED, INC. c/o GRACE MEDICAL GROUP, INC., 8500 Wolf Lake Drive, Suite 110, Memphis, TN 38133, or by fax to 901-386-0950 or email at [email protected].
Recall details
- Recall number
- Z-1521-2021
- Status
- Open, Classified
- Initiated by the company
- Dec 7, 2020
- Recalling firm
- Oto Med Inc
- Quantity
- 415 devices
- Product code
- MXP
- Distribution
- Worldwide distribution - U.S. Nationwide distribution including in the states of MN, IL, PA, WI, GA, SD, FL, AR, MA, AZ, TX, MO, NC, CA, UT, DE, OR, IN, CO, AL, OH, NY, WA, DC, TN, NH, VT, VA, SC, MI, WI, MS, IA, NJ, MD, KS, CT, LA, ME, KY, HI, TN, RI, NV, ID and the countries of Argentina, Brazil, Colombia, Chile, Hong Kong, South Africa, United Arab Emirates, Taiwan, Canada, India.
Product description
Grace Medical REF S-4500 *** spineX BONE DUST COLLECTOR *** 2 stainless steel filters - Product Usage: intended to be inserted into a suction tube during the early stages of otologic surgery to collect bone particles for future use.
Code information (models, lots, serial numbers)
UDI Code: 00850000126077 Lot Numbers: 019120 G(2) 0712119 G(2) 0912118 G(2) 1012119 G(2)
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- spineX BONE DUST COLLECTOR S-4500Oto-Med, Inc. · Apparatus, Suction, Operating-Room, Wall Vacuum Powered1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 87577).
- Z-1519-2021 · Open, ClassifiedGrace Medical REF S-2500 *** SHEEHY BONE DUST COLLECTOR *** Stainless steel filter - Product Usage: intended to be inserted into a suction tube during the early stages of otologic…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page