Trinity Biotech Captia Measles IgM Kit, Catalog No. 2326060… recallZ-1503-2021 · Clark Laboratories, Inc. · initiated Mar 17, 2021

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Clark Laboratories, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Positive Control may run high out of its assigned ISR range and result in an invalid run and necessitate retesting

FDA root cause category: Under Investigation by firm

What the company is doing

Trinity Biotech issued Product Notification Letter on March 17, 2021 stating reason for recall, health risk and action to take: Please note, if a laboratorys test resulted in a valid run, such as all kit Controls and Calibrator met the Quality Control Criteria as stated in the package insert, the patient results can be reported and a re-evaluation of patient results is not necessary. Customers are instructed to discontinue the use of any Measles IgM Lot #2326060-060 kits that they may have in their inventory. Please complete the attached fax back form informing us of how many of these kits you and your customers have in your inventory that will require replacement. A thorough investigation into the root cause of this issue has been initiated and any corrective/preventive actions will be implemented as appropriate. If you have any questions or comments arising from this customer communication, the Trinity Biotech Help Desk is ready to answer your questions (800-325-3424, option 2).

Recall details

Recall number
Z-1503-2021
Status
Terminated
Initiated by the company
Mar 17, 2021
Recalling firm
Clark Laboratories, Inc.
Quantity
80 kits
Product code
PCL
FDA submissions
K140455
Distribution
Worldwide distribution - US Nationwide distribution and the country of Ireland.

Product description

Trinity Biotech Captia Measles IgM Kit, Catalog No. 2326060 - Product Usage: intended for the qualitative detection of Measles igM antibodies in human serum of patients suspected of measles (rubeola) infection.

Code information (models, lots, serial numbers)

Kit Lot: 2326060-060 Expiration Date: 2021-06-28 (UDI): 05391516743655

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates