Skip to content

Direct Inject recallZ-1436-2021 · Stryker Leibinger GmbH & Co. KG · initiated Apr 1, 2021

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Stryker Leibinger GmbH & Co. KG or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Inability for the user to inject the paste from the syringe into the target location.

FDA root cause category: Under Investigation by firm

What the company is doing

On 04/01/2021, Stryker issued an Urgent Medical Device Removal notice via letter informing customers that users have reported the inability to push the paste through the Mixer-Cannula which prohibits the injection into a bone void.

Recall details

Recall number
Z-1436-2021
Status
Terminated
Initiated by the company
Apr 1, 2021
Recalling firm
Stryker Leibinger GmbH & Co. KG
Quantity
321 units
Product code
GXP
FDA submissions
K143661
Distribution
Worldwide distribution - US Nationwide distribution and the countries of Germany, Italy, France, Australia, Japan.

Product description

DirectInject consists of a sterile dual paste system which is calcium phosphate based. Upon injection through the Mixer-Cannula, the two pastes form a cement paste which is moldable. The injected cement paste will harden under normal body conditions to form hydroxyapatite.

Code information (models, lots, serial numbers)

Lot: DI20311,DI20307  Product Code (UDI): 07613327123265

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

2 devices cleared under K143661 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page