Colpassist Vaginal Positioning Device - Product Usage:… recallZ-1490-2021 · Boston Scientific Corporation · initiated Apr 15, 2021
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Boston Scientific Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Potential for pinholes (sterile barrier breach) on the edge of the pouch to potentially render the device non-sterile, and may result in a post-operative infection
FDA root cause category: Packaging process control
What the company is doing
Boston Scientific issued Urgent Medical Device Product Removal on 4/15/21 via FedX. Letter states reason for recall, health risk and action to take; Individually packaged Upsylon Y Mesh (UPN M0068318200) is not impacted by this removal. This removal affects only the UPNs and batches listed below and in Attachment 1. No other UPNs or batches are impacted by this removal. Please cease further distribution or use of any remaining product affected by this removal immediately. If you identify any product from the affected lots/batches within your inventory, please segregate the product immediately and return it to BSC in accordance with the enclosed instructions and indicate on your Reply Verification Tracking Form the quantity of units from each lot/batch that you will be returning. BSC is notifying regulatory authorities of this removal as required. Please read carefully through the removal instructions included with this notification. Your local Sales Representative can answer any questions that you may have regarding this notification.
Recall details
- Recall number
- Z-1490-2021
- Status
- Open, Classified
- Initiated by the company
- Apr 15, 2021
- Recalling firm
- Boston Scientific Corporation
- Quantity
- 4,549 WW
- Product code
- KOH
- Distribution
- Worldwide distribution - US Nationwide distribution and the countries of Algeria, Brazil, Colombia, Estonia, Germany, Hong Kong, Australia, Canada, Costa Rica, Finland, Great Britain, Italy, Austria, Chile, Czech Republic, France, Greece, Jordan, Belgium, Lebanon, Mauritius, Mexico, Morocco, Namibia, Netherlands, New Zealand, Norway, Panama, Poland, Portugal, Puerto Rico, Serbia, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan.
Product description
Colpassist Vaginal Positioning Device - Product Usage: intended for use in general gynecological surgery to assist in the positioning and manipulation of the vagina during the procedure. Material No. M0068318210
Code information (models, lots, serial numbers)
All unexpired lots: C003699 C003715 C003735 C003750 C003775 C003804 C003813 C003842 C003853 C003887 C003905 C003936 C003950 C003960 C003966 C003986 C003997 C004028 C004041 C004078 C004091 C004107 C004117 C004128 C004136 C004160 C004186 C004206 C004221 GTIN 08714729848097
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Colpassist M0068318210Boston Scientific Corporation · INSTRUMENT, MANUAL, GENERAL OBSTETRIC-GYNECOLOGIC1 recall2 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 87709).
- Z-1491-2021 · Open, ClassifiedUpsylonTM Y Mesh Kit with Colpassist Vaginal Positioning Device - Product Usage: intended for use as a bridging material for sacrocolposuspension/sacrocolpopexy (laparotomy, laparo…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page