Skip to content

AVID TruCustom recallZ-1787-2021 · Avid Medical, Inc. · initiated Apr 9, 2021

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Avid Medical, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Medical convenience kits are being recalled due to a potential breach in the package integrity of a component, the BD ChloraPrep 3mL applicator. Under certain circumstances the applicator may grow Aspergillus penicillioides, which may result in a breach of the package integrity. A breach of package integrity may contaminate other components in AVID Medical kits.

FDA root cause category: Environmental control

What the company is doing

A Recall notification letter titled, "URGENT MEDICAL DEVICE CORRECTION" was sent to consignees on 04/09/2021 via a email. End-users were notified on 4/13/21. The letter instructs the consignee to immediately examine all inventory locations and discontinue use of all Halyard or AVID branded Procedure Trays subject to the field action. Discard all affected Procedure Trays found in inventory. In addition, if the consignees has further distributed this product, the notice instructs the consignee to please identify their customers and notify them at once of this product correction and requirements. The notification to the consignee's customers may be enhanced by including a copy of this URGENT MEDICAL DEVICE CORRECTION Notice. The consignee is also asked to please complete the attached Field Action Response Form (page 3) immediately, even if they no longer have any of the recalled lot numbers left in inventory. The completed form is to be emailed to: [email protected]. If the facility has identified affected product in the consignee's inventory, the recall coordinator will reach back to their facility and provide the consignee with paperwork for documenting destruction of the recalled product.

Recall details

Recall number
Z-1787-2021
Status
Terminated
Initiated by the company
Apr 9, 2021
Posted by the FDA
Jun 4, 2021
Recalling firm
Avid Medical, Inc.
Quantity
560 kits
Product code
LRO
Distribution
Distributed US nationwide to TX, AZ, IL, MS, LA, MO, and TN

Product description

DIALYSIS ACCESS Medical convenience kit, Avid Medical Part Number: VMED004-03.

Code information (models, lots, serial numbers)

Part Number:VMED004-03; Package GTIN: 20809160238247; Lot Number: 1409970 (Expiration Date: 04/26/2022), 1412734 (Expiration Date: 10/01/2021), 1413146 (Expiration Date: 10/01/2021), and 1413395 (Expiration Date: 04/25/2022)

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 87720).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page