Skip to content

Alaris Pump Module, Model 8100, serviced/repaired by… recallZ-1543-2021 · Step-Har Medical LLC · initiated Feb 26, 2021

Class I Terminated All recalls posted in May 2021

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Step-Har Medical LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Bezel repair parts used, not by the original manufacturer, to service and repair infusion pump modules, may over time experience cracking or separation of the posts from the bezel. The separation of one or more bezel repair part posts may result in free flow, over infusion, under infusion or interruption of infusion.

FDA root cause category: Component design/selection

What the company is doing

On 02/26/2021, the one customer was notified of the recall via phone call and email. On 04/15/21, a Voluntary Field Safety Corrective Action notice were mailed. The firm asked their customer to respond so the bezel repair part could be replaced. Customers can contact the recalling firm at: 657-267-0054 or [email protected]

Recall details

Recall number
Z-1543-2021
Recall class
Class I (most serious)
Status
Terminated
Initiated by the company
Feb 26, 2021
Posted by the FDA
May 10, 2021
Recalling firm
Step-Har Medical LLC
Quantity
2
Product code
FRN
FDA submissions
K133532
Distribution
IA

Product description

Alaris Pump Module, Model 8100, serviced/repaired by Step-Har Medical with an impacted bezel repair part

Code information (models, lots, serial numbers)

Serial Numbers: 12747983 and 9919591

Other devices under the same FDA submission

4 devices cleared under K133532 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page