Skip to content

FREESTYLE LIBRE 2 GLUCOSE MONITORING SYSTEM: Item number:… recallZ-1931-2021 · Cardinal Health Inc. · initiated Mar 16, 2021

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Cardinal Health Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Due to shipping delays the devices were exposed to temperature excursions of 31.9F, intermittently over a 2 hour and 15 minute period of time which may cause inaccurate results.

FDA root cause category: Environmental control

What the company is doing

On 03/16/2021, Cardinal Health issued an Urgent Medical Device Removal notice to customer via letter notifying them a recall is being initiated as a precautionary measure as a result of exposure to temperatures outside of the products labeling requirements. Actions Required 1. INSPECT your inventory for the affected product. 2. SEGREGATE and QUARANTINE all affected product. 3. RETURN the enclosed acknowledgment form via fax to 1-817-868-5362 or email to [email protected], even if you do not have any product on hand. 4. CONTACT Inmar to arrange for return and credit of any affected product: 1-855-696-2074 (9:00 AM to 5:00 PM, Monday thru Friday)

Recall details

Recall number
Z-1931-2021
Status
Open, Classified
Initiated by the company
Mar 16, 2021
Recalling firm
Cardinal Health Inc.
Quantity
2,115 kits
Product code
QLG
FDA submissions
K193371
Distribution
US Nationwide distribution in the states of FL, GA, SC.

Product description

FREESTYLE LIBRE 2 GLUCOSE MONITORING SYSTEM: Item number: 5658877

Code information (models, lots, serial numbers)

NDC/UPC: 57599080000 All lots numbers shipped from 2/23/201 to 3/10/2021

Other devices under the same FDA submission

1 device cleared under K193371 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 87787).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page