NeuroTherm NT2000iX Radiofrequency (RF) Generator, Model… recallZ-1549-2021 · Abbott Laboratories Inc. (St Jude Medical) · initiated Apr 22, 2021
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Abbott Laboratories Inc. (St Jude Medical) or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Potential capacitor failure which typically manifests as in inability to provide output on one or more channels. In some instances, users reported a charring/smoky odor or emission.
FDA root cause category: Process change control
What the company is doing
The firm issued a medical device correction/service notice beginning on 04/22/2021. The notice discusses device failure and how to identify it. Should the failure present, then firm is requesting return of the device for servicing.
Recall details
- Recall number
- Z-1549-2021
- Status
- Terminated
- Initiated by the company
- Apr 22, 2021
- Recalling firm
- Abbott Laboratories Inc. (St Jude Medical)
- Quantity
- 318 units
- Product code
- GXD
- FDA submissions
- K111576
- Distribution
- US and AUSTRALIA, EGYPT, SOUTH AFRICA, BELGIUM, COLOMBIA, UNITED KINGDOM, COSTA RICA, ISRAEL, SOUTH KOREA, SINGAPORE, TAIWAN
Product description
NeuroTherm NT2000iX Radiofrequency (RF) Generator, Model Number RFG-NT-2000
Code information (models, lots, serial numbers)
Show the full code information (3k characters)
UDI 05415067022417: Serial Numbers: 18675078, 18709738, 18097502, 18023105, 18739462, 18650501, 18493702, 18718469, 18666645, 18634334, 18666647, 18427317, 18742999, 18709732, 18222185, 18222187, 18240730, 18201200, 18743003, 18343825, 18088294, 18201198, 18222190, 18240731, 17991994, 18132016, 18500164, 18522935, 18659497, 18750318, 18854584, 18718473, 18873320, 18240735, 18102325, 18648497, 18666643, 18666646, 18657521, 17991999, 18313159, 18614034, 18721060, 18493705, 18493709, 18610701, 18302046, 18192751, 18132012, 18222188, 18240728, 18718472, 18500167, 18132008, 18493711, 18313153, 17905956, 18060073, 18132011, 18102323, 18014268, 18014258, 18343824, 18240733, 18180086, 18675069, 18060017, 18060067, 18060071, 17964784, 18427318, 18648518, 18675068, 18648648, 18192752, 18709733, 18709734, 18302042, 18265068, 18659496, 18192753, 18192756, 18222183, 18522719, 18705743, 18522937, 18644839, 18873319, 18102324, 18088297, 18659489, 18706571, 18192754, 18622842, 18240729, 18180085, 18659495, 18187525, 18192750, 18115250, 18714394, 18666648, 18070518, 18072514, 18180082, 18634336, 18699641, 18854582, 18522709, 18704516, 18493701, 18716594, 18709741, 18873321, 18302048, 18644838, 18699645, 18102337, 18555754, 18743000, 18690921, 18180088, 18657520, 18522715, 18343821, 18201195, 18648650, 18522714, 18265071, 18709739, 18709735, 18023104, 18313165, 18070527, 18648652, 18650503, 18265072, 18666650, 18634337, 18313157, 18533473, 18648647, 18705742, 18644824, 18493707, 18699643, 18265070, 18343819, 17905963, 18070525, 18750319, 18343820, 18343823, 18650500, 17991997, 18701492, 18742994, 18014262, 18709742, 18854581, 18427321, 18659494, 18659493, 18644840, 18500166, 18533504, 18180083, 18427319, 18648008, 18648578, 18690926, 18657519, 18014259, 18014254, 18132092, 18302043, 18132014, 18003308, 18003311, 18003313, 18014255, 18014269, 18659490, 18047384, 18088295, 18014267, 18038591, 18023107, 18003310, 18014261, 18648619, 18648653, 18648655, 18102335, 18265069, 18854583, 18222191, 18014270, 18014271, 18690920, 18302047, 18696583, 18522716, 18427320, 18265066, 18302041, 18699642, 18192749, 18192755, 18675070, 18675076, 17860513, 17942355, 17959675, 17959676, 17959677, 17981234, 17981235, 18003314, 18003325, 18014272, 17860514, 17860515, 17905959, 17905962, 18187528, 18892887, 17905965, 17863602, 18302049, 18500165, 17856739, 18115251, 17856738, 17903459, 18115245, 18115246, 18126612, 18126613, 18500161, 17865145, 17856737, 18187529, 17981236, 17981233, 18132007, 18699646, 18180087, 18699640
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- NT2000iX™ RFG-NT-2000St. Jude Medical, Inc. · GENERATOR, LESION, RADIOFREQUENCY2 recalls310 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page