Skip to content

IntelliVue MX40 Patient Monitor, Model Number: 865350 (US)… recallZ-1671-2021 · Philips North America Llc · initiated Feb 2, 2021

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Philips North America Llc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Manufacturing failed to follow process steps resulting in the distribution of potentially defective devices

FDA root cause category: Employee error

What the company is doing

Philips Healthcare conducted site visits on 2/02/21 by Field Service Engineer and provided the customer with RMA to return affected devices. Devices retuned affected device on RMA. Device Replacement provided.

Recall details

Recall number
Z-1671-2021
Status
Terminated
Initiated by the company
Feb 2, 2021
Recalling firm
Philips North America Llc
Quantity
3 units
Product code
DSI
FDA submissions
K113125
Distribution
Worldwide distribution - US Nationwide distribution in the states of IN, MN and the country of Italy.

Product description

IntelliVue MX40 Patient Monitor, Model Number: 865350 (US) 865352 (OUS) Software version number B.06.59 - Product Usage: Intended for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adults and pediatrics in hospital environments and during transport inside hospitals.

Code information (models, lots, serial numbers)

Serial numbers: US09699788 (US) US156J6219 (US) US156J3658 (OUS)   UDI:  (01)00884838082236 (21)US09699788 (01)00884838082236 (21)US156J6219 (01)00884838030350 (21)US156J3658

Devices named in this recall

2 devices in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page