Skip to content

T25 Cannulated Screwdrivers (CSD 025) packaged in Guide… recallZ-1698-2021 · Flower Orthopedics Corporation · initiated Apr 15, 2021

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Flower Orthopedics Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Cannulated Screwdrivers may break at the driver tip along the shaft, near the handle extending surgery time and additional x-ray to identify the fragments from the breakages

FDA root cause category: Nonconforming Material/Component

What the company is doing

Flower Orthopedic issued Urgent Recall Information via email on 4/15/21. Letter states reason for recall, health risk and action to take: Check your inventory for the product listed below and quarantine immediately. Next, please complete and return the attached Acknowledgement and Receipt Form as soon as possible. Flower Orthopedics is initiating shipment of replacement product for delivery Friday April 16. Contact information for questions: Monday through Friday, 8:30 AM to 5:00 PM, EDST 877-778-8587 or [email protected] "Acknowledgement and Product Replacement Forms (see attached)

Recall details

Recall number
Z-1698-2021
Status
Terminated
Initiated by the company
Apr 15, 2021
Recalling firm
Flower Orthopedics Corporation
Quantity
47 units
Product code
HXX
Distribution
US Nationwide distribution in the states of AZ, CO, FL, GA, IL, IN, PA, NC, VA.

Product description

T25 Cannulated Screwdrivers (CSD 025) packaged in Guide Wire Kits (GWK 300) - Product Usage: Intended to be used to insert screws with T25 driving features, which includes screws with diameters of 5.5mm, 6.5mm, and 7.3mm Product Code: GWK 300

Code information (models, lots, serial numbers)

Lot Numbers: 2003320898, 1810220041, 1802220008, 1802220005, 1711220035  UDI: 00840118110853

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page