Continuous CT Fluoroscopy DVI Ceiling Mount CCT Arm DVI… recallZ-1988-2021 · Philips North America Llc · initiated May 1, 2021
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Philips North America Llc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Significant wear of the Monitor Bracket Assembly shaft goes unrecognized could potentially lead to the Monitor Bracket Assembly detaching from the Suspension Arm and result in potential hazard for the monitor to fall while remaining attached to the cables resulting in patient, operator, bystander, or service person to serious injury.
FDA root cause category: Device Design
What the company is doing
Philips issued URGENT- Field Safety Notice Medical Device Correction letter on 5/1/21 FSN 72800751 and FSN 88200530 . Letter states reason for recall, health risk and action to take: " Be aware of this hazard when the safety clip is missing from the joint assembly between the Monitor Bracket Assembly and the Ceiling Suspension Arm. " Inspect the joint area of their CCT DVI Ceiling Mount Arm (the circled area in picture at theright) and observe that the safety clip is installed (as depicted in pictures below). " When this hazard occurs, the monitor may become difficult to rotate relative to the Suspension Arm. When these symptoms are recognized, in order to avoid the occurrence of a hazardous situation, take the following actions: stop using the suspension monitor clinically move the monitor arm away from any person call Philips Service for repairs. Please acknowledge the receipt of this notice by completing, signing, and returning the form attached at the end of this communication to the indicated email address. Philips will inspect systems via Field Change Order FCO 88200530. Systems failing the inspection will be corrected via Field Change Order FCO 882531 For North America, contact the Customer Care Solutions Center (1-800-722-9377, follow the prompts).
Recall details
- Recall number
- Z-1988-2021
- Status
- Open, Classified
- Initiated by the company
- May 1, 2021
- Recalling firm
- Philips North America Llc
- Quantity
- 1,301 Monitor Bracket Assemblies total
- Product code
- KPS
- FDA submissions
- K123599
- Distribution
- Worldwide distribution - US Nationwide and the countries of Canada, Mexico, AE, AR, AT, AU, BD, BE, BR, CH, CL, CN, CO, CU, CY, CZ, DE, DK, DZ, EE, EG, ES, FI, FR, GB, GE, GI, GL, GR, HK, HU, ID, IE, IL, IN, IQ, IR, IT, JO, JP, KE, KG, KH, KR, LA, LB, LK, LT, LY, MM, MN, MU, MV, MY, NI, NL, NO, NZ, OM, PA, PF, PH, PK, PL, PS, PT, QA, RE, RU, RW, SA, SE, SG , SI, SK, TH, TN, TR, UZ, VN, XK, YE, ZA, ZW.
Product description
Continuous CT Fluoroscopy DVI Ceiling Mount CCT Arm DVI Option - Monitor 882446 - Vereos PET/CT
Code information (models, lots, serial numbers)
Model/ Serial Number/UDI: N/A 882446 900115 (01)00884838059559(21)900115
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Vereos PET/CTPhilips Medical Systems Nederland B.V. · System, tomography, computed, emission1 recall1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 87974).
- Z-1982-2021 · Open, ClassifiedContinuous CT Fluoroscopy DVI Ceiling Mount CCT Arm DVI Option - Monitor Bracket Assembly for use with CT and PET/CT systems. Brilliance 64, Brilliance 40, Ingenuity Flex 728…
- Z-1983-2021 · Open, ClassifiedContinuous CT Fluoroscopy DVI Ceiling Mount CCT Arm DVI Option - Monitor Bracket Assembly for use with CT and PET/CT systems. BB (Big Bore) Oncology, BB (Big Bore) Radiology:…
- Z-1984-2021 · Open, ClassifiedContinuous CT Fluoroscopy DVI Ceiling Mount CCT Arm DVI Option - Monitor Ingenuity Core, Ingenuity Core 128, Ingenuity CT 728321 - Ingenuity Core 728323 - Ingenuity Core 12…
- Z-1985-2021 · Open, ClassifiedContinuous CT Fluoroscopy DVI Ceiling Mount CCT Arm DVI Option - Monitor 728332 - IQon Spectral CT : IQon Spectral CT
- Z-1986-2021 · Open, ClassifiedContinuous CT Fluoroscopy DVI Ceiling Mount CCT Arm DVI Option - Monitor 728333 - Spectral CT:
- Z-1987-2021 · Open, ClassifiedContinuous CT Fluoroscopy DVI Ceiling Mount CCT Arm DVI Option - Monitor 882442/882456 - Ingenuity TF PET/CT
- Z-1990-2021 · Open, ClassifiedContinuous CT Fluoroscopy DVI Ceiling Mount CCT Arm DVI Option - Monitor 728306 - Brilliance iCT 728311 - Brilliance iCT SP
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page