The ADVIA Chemistry XPT Chemistry System, Siemens Material… recallZ-2014-2021 · Siemens Healthcare Diagnostics, Inc. · initiated May 18, 2021
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Siemens Healthcare Diagnostics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
ADVIA Chemistry System and Software Versions (V1.4 SMN 11314625 and V1.4 China SMN 11316885) may not be aligned with customer expectations for the current default mixer failure detection setting of 2; an erroneously elevated or depressed patient result to be reported without alert on the test result
FDA root cause category: Software design
What the company is doing
Siemes issued an Urgent Medical Device Correction (UMDC) was mailed to all US affected customers by FedEx beginning on 05/18/2021. An Urgent Field Safety Notice (UFSN) was issued to all OUS Siemens sites with affected customers electronically for distribution and regional reporting to their local competent authorities beginning on 05/18/2021. Letter states reason for recall, health risk, and action to take: Until the next version of instrument software is available Siemens recommends changing the software behavior from the current default of 2: Warning+STOP to 1: Warning. The behavior for the Warning setting is to display the test result as ERROR with a Safety(S) flag and generate a yellow warning in the Operators event log for each test result encountering a mix error. The instrument will continue to process samples. The flagged test must be repeated to obtain a valid result. If you have any questions, please contact your Siemens Healthineers Customer Care Center or your local Siemens Healthineers technical support representative. The ADVIA Chemistry XPT version 1.6 software will be released in June 2021
Recall details
- Recall number
- Z-2014-2021
- Status
- Terminated
- Initiated by the company
- May 18, 2021
- Recalling firm
- Siemens Healthcare Diagnostics, Inc.
- Quantity
- 842 units
- Product code
- JGS
- FDA submissions
- K990346
- Distribution
- Worldwide distribution - US Nationwide distribution in the states of CA, CO, CT, FL, MA, MD, MI, MN, NY, OH, PA, TX and the countries of Australia, Bahrain, Brazil, Canada, Chile, Croatia, Denmark, Egypt, Estonia, Finland, France, Georgia, Germany, Greece, Hong Kong, Hungary, India, Italy, Malaysia, Netherlands, New Zealand, Norway, P.R. China, Portugal, Republic Korea, Saudi Arabia, Slovakia, Spain, Sweden, Taiwan, Thailand, Turkey, United Kingdom, Vietnam.
Product description
The ADVIA Chemistry XPT Chemistry System, Siemens Material Number (SMN) 10723034, is an automated, clinical chemistry analyzer that runs tests on serum, plasma, urine, or cerebral spinal fluid in random access and batch modes. Tests performed using this system are intended for in vitro diagnostic use.
Code information (models, lots, serial numbers)
Software version: 1.4 or lower UDI - 00630414595467
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- ADVIA® Chemistry XPT System 10723034Siemens Healthcare Diagnostics Inc. · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE1 recall174 reports
Other devices under the same FDA submission
1 device cleared under K990346 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- ADVIA® 1800 Analyzer 10315183Siemens Healthcare Diagnostics Inc. · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE150 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page