NeuMoDx Cartridge, For In Vitro Diagnostic Use on the… recallZ-1978-2021 · NeuMoDx Molecular Inc · initiated May 4, 2021

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from NeuMoDx Molecular Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

There is a potential for false positive results when certain lots of cartridges are used in conjunction with specific assays.

FDA root cause category: Material/Component Contamination

What the company is doing

URGENT MEDICAL DEVICE CORRECTION notification letters dated 5/4/21 were sent to customers. Actions Required by Customers 1. IMMEDIATELY discontinue use of the product and discard remaining inventory in accordance with your national and local safety and environmental regulations. Please contact QIAGEN Technical Service for a free-of-charge replacement of scrapped inventory. 2. Share this Medical Device Correction Notice with all users of the NeuMoDxTM Cartridges in your facility to ensure they are aware of this notice. 3. If you have already used NeuMoDxTM Cartridges from any of these lots in combination with the NeuMoDxTM SARS-CoV-2 Assay or the NeuMoDx Flu-A-B/RSV/SARS-CoV-2 Assay, please review all SARS-CoV-2 positive results to exclude erroneous diagnosis and treatment, except in those cases where alternative confirmation was obtained. Note: results for targets other than SARS-CoV-2 are not impacted by this medical device correction. 4. Complete the attached response form and return to [email protected] to confirm receipt of this notification. We sincerely apologize for any inconvenience this may have caused and thank you in advance for your patience and cooperation. If you have any questions regarding this notice, please contact QIAGEN Technical Services at 1-800-362-7737 or [email protected]

Recall details

Recall number
Z-1978-2021
Status
Terminated
Initiated by the company
May 4, 2021
Recalling firm
NeuMoDx Molecular Inc
Quantity
620 cases (29,760 cartridges total)
Product code
QJR
Distribution
Worldwide distribution.

Product description

NeuMoDx Cartridge, For In Vitro Diagnostic Use on the NeuMoDx 288 and NeuMoDx 96 Molecular Systems, Model Number 100100

Code information (models, lots, serial numbers)

GTIN: 10814278020274;  Lot Numbers: 106629 106630 106631 106632

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.